Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 62
System Requirements
System Recommendations
(for Optimal Usage)
Screen Resolution
1024 x 768
≥ 1280 x 1024
Internet Bandwidth
56 kB/sec
≥ 128 kB/sec
(DSL, cable modem)
PC
800 MHz Pentium
processor, 128 MB RAM
N/A
Internet Browser
MS Internet Explorer 5.5
≥ MS Internet Explorer 5.5
- or -
≥ Mozilla 1.8 (Firefox ≥ 2.0)
Acrobat Reader
Version 6.1
Version 6.1 or higher
Communication
Channel
Fax (G3) or e-mail
Fax (G3), e-mail or mobile phone
Table 18: System Requirements / Recommendations
Additionally, the attending physician may register to be informed of the occurrence of an Event
Triggered Message through email or SMS (i.e., mobile phone) with a brief text message. If registered
for Internet availability, the patient’s detailed implant data can then be viewed by logging onto the Home
Monitoring website.
5.18.4 Types of Report Transmissions
When the Home Monitoring function is activated, the transmission of a report (Cardio Report) from the
implant can be triggered as follows:
• Trend report—the time period (daily) initiates the report
• Event report—the pulse generator detects certain events, which initiate a report
5.18.4.1 Trend Report
The time of the report transmission is programmable. For periodic messages, the time can be set
anywhere between 0:00 and 23:59 hours. It is recommended to select a time between 0:00 and 4:00.
The length of the time interval (monitoring interval) is preset to “daily”. For each monitoring interval, a
data set is generated in the implant and the transmission is initiated at the designated time.
5.18.4.2 Event Report
When certain cardiac and technical events are detected by the implant, a report transmission is
automatically triggered. This is described as an “event message” as part of the daily transmission.
The following clinical and technical events initiate a Home Monitoring message transmission:
• Atrial Lead Check < 100+/-50 Ohm and > 2500+/-500 Ohm
• Right/Left Ventricular Lead Check < 100+/-50 Ohm and > 2500+/-500
• RVCC/LVCC Disabled
• ERI detected
• Atrial Capture Control Disabled
• High Ventricular rate
• Atrial tachyarrhythmia persisting beyond a programmable time limit or
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...