Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 17
The triggered pacing mode fixes the AV delay to 180 ms and does not provide a safety AV delay.
Pacing modes with triggered response may be indicated in the presence of interference signals to
prevent inappropriate pulse inhibition. They may also have diagnostic application for ECG identification
of sense events as an alternative to marker signals. Triggered pacing may also be used for
hemodynamic as well as electrophysiologic studies and for termination of tachycardias by non-invasive
triggering of pulse generator pulses with chest wall stimuli generated by an external pulse generator.
CAUTION
Programmed to Triggered Modes
When programmed to triggered
modes, pacing rates up to the programmed upper limit may occur in
the presence of either muscle or external interference.
Triggered Modes
While the triggered modes (DDT, VVT, and AAT) can
be programmed permanently, the use of these modes is intended as
a temporary setting in situations where maintaining the programming
head in place would be impossible or impractical (i.e., during exercise
testing or extended Holter monitoring) or as a short term solution to
pulse generator inhibition by extracardiac interference. To avoid the
potential for early battery depletion, it is important that the triggered
modes are not used for long term therapy, and that the pulse generator
is returned to a non-triggered permanent program.
5.2 Rate Related Functions
The availability of parameters and parameter values is determined by the software used for
programming/ interrogating the pulse generator.
5.2.1 Basic Rate
) sets the lower pacing rate for the pacemaker and may be
programmed from 30 bpm to 200 bpm. Evia HF / HF-T will allow pacing lower than the programmed
rate when the parameter Hysteresis is enabled. The Hysteresis parameter is found under Basic rate /
Night rate.
Programming conflicts for Basic rate occur when Atrial and/or Ventricular Capture Control is
programmed ON. When capture control is programmed ON, the basic rate is limited to 100 bpm.
Figure 2: Basic Rate/Night Rate Screen with Default Settings
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...