Chapter 12
Explantation
Evia HF / HF-T Technical Manual
PAGE 114
Source
Cause
Possible Effect
Cardiac perforation
The above plus cardiac tamponade; muscle or
nerve stimulation.
Myocardial irritability at time of
insertion, e.g., from an acute
myocardial infarction
Fibrillation
Patient
Threshold Elevation
Loss of capture and/or sensing.
Normal medical complication
Infection
Body rejection phenomena
Fluid accumulation; migration; erosion.
Misc.
Unipolar pacing systems
Inhibition of pulse generator due to sensing of
skeletal muscle activity.
Physician preference
Upgrade to bipolar, dual chamber, rate-adaptive
pulse generator, etc.
Introducer caused
Air embolism or pneumothorax.
Table 25: Common Reasons to Explant a Pulse Generator
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...