Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 55
atrial event will be detected either due to the emerging intrinsic rhythm or due to retrograde conducted
events caused by ventricular paces. The detection of sensed atrial activity is used to discriminate
between atrial capture and non-capture. The atrial capture control is performed in four steps:
1. Setup-Phase: The device monitors the rate and rhythm condition and determines the actual rate
in the atrium immediately before it starts the threshold search. Automatic measurements are
allowed if the atrial and ventricular rate is below 110 ppm and no mode switching is active. If
these conditions are met, the activation of the capture control algorithm causes a mode switch
to DDI with an atrial overdrive pacing of +20 % of the actual determined rate. The Ap-Vp interval
will be programmed to 50 ms, to avoid retrograde conduction from the ventricle.
2. Threshold search: The threshold is determined by decreasing the amplitude stepwise at a
programmed pulse duration until loss of capture occurs. Loss of capture for one test amplitude
is declared if in a test window of five cardiac cycles (5 Ap-Vp intervals) two or more intrinsic
atrial events are sensed which indicates unsuccessful pacing.
3. Confirmation phase: The pacing threshold is considered to be confirmed if capture is determined
with the first step and loss of capture is confirmed with the second step.
4. Amplitude adjustment: The pacing amplitude is defined by adding the programmed safety
margin to the determined threshold.
Parameters associated with Atrial Capture Control are shown in
Parameter
Range
Default
Capture Control
ON, ATM, OFF
ON
Minimum Amplitude
0.5...(0.1)...4.8 V
1.0 V
Threshold Test Start
2.4, 3.0, 3.6, 4.2, 4.8 V
3.0 V
Safety Margin
0.3...(0.1)...1.2 V
1.0 V
Search Type
Time of day, Interval
Time of day
Time of Day
24 hour clock
2:00 A.M.
Interval
0.1, 0.3, 1, 3, 6, 12, 24 hr.
24 hr.
Table 15: Atrial Capture Control Parameters
Capture Control
Figure 34: Atrial Capture Control Screen
Содержание Evia HF
Страница 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Страница 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Страница 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Страница 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Страница 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Страница 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Страница 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Страница 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Страница 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Страница 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...