![BIOTRONIK Evia DR Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/biotronik/evia-dr/evia-dr_technical-manual_2756148083.webp)
Evia Technical Manual 75
The following functions are available for tailoring the motion
based rate adaptation to the individual patient except for the
maximum closed loop rate, which is relevant to CLS.
7.7.1 Sensor Gain
The sensor gain defines the slope of the linear function between
exertion and pacing rate. It designates a factor by which the
electric signal of the sensor is amplified prior to the signal
processing stages. The programmable amplification permits
adaptation of the individually programmed sensor gain to the
desired rate response. The optimum setting is achieved when
the desired maximum pacing rate during exertion is reached
during maximum exercise levels. The rate increase, rate
decrease and maximum sensor rate settings must be checked
for their suitability with respect to the individual patient before
adjusting the sensor gain.
If the sensor-driven rate is not sufficient at high levels of exertion
the sensor gain setting should be increased. The sensor gain
should be reduced if high pacing rates are obtained at low levels
of exertion.
Figure 13. Influence of sensor gain on the rate response.
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......