100 Evia Technical Manual
Table 25: Dual Node D Paroxysmal AV Block and
Vasovagal Syncope
Standard
Program
Dual Node
D
Paroxysmal
AV Block
Vasovagal
Syncope
Valid for: DR(-T)
DR(-T)
DR(-T)
Parameters
Mode DDD-R
DDD-CLS
DDD-CLS
Rate Hysteresis
(ppm)
OFF ---
---
CLS Vp required
---
No
No
CLS Resting
Rate Control
--- +20 OFF
AV Hysteresis
OFF
IOPT
IOPT
Holter settings
High atrial rate
AT
ModeSw
OFF
High ven. Rate
ON
ON
OFF
Pre-trigger
recording (%)
75 75 ---
IEGM signal
Filtered
Filtered
---
HAR limit (bpm)
200
200
---
HVR limit (bpm)
180
160
---
HVR counter
8
8
---