![BIOTRONIK Evia DR Скачать руководство пользователя страница 103](http://html1.mh-extra.com/html/biotronik/evia-dr/evia-dr_technical-manual_2756148103.webp)
Evia Technical Manual 95
A SA/CV sequence will be postponed when the following events
occur:
•
Mode switch is ongoing: If Mode switch is ongoing when
an SA/CV is scheduled, the SA/CV will be postponed
until reversion back to the programmed mode.
•
The presence of a magnet is detected: A scheduled
SA/CV is postponed when a magnet (programmer wand)
is detected. After removal of the magnet, the SA/CV is
started.
•
The ventricular rate is higher than 110 bpm: A scheduled
SA/CV is postponed when the ventricular rate is higher
than 100 bpm. When the ventricular rate drops below
100 bpm, the SA/CV is started.
The continuous capture control (CV) will be suspended and the
ventricular pacing amplitude is set to high output when the
following events occur:
•
The ventricular rate is higher than 110 bpm: CV is
suspended while the ventricular rate is higher than
110 bpm. When the ventricular rate drops below
110 bpm, CV is resumed.
•
Mode Switch: CV is suspended and the device is set to
high output until reversion back to the programmed
mode.
The VCC feature will be automatically turned OFF when the
following events occur:
•
The initial SA/CV sequence after activating VCC failed
•
Three subsequent and consecutive failed SA attempts
•
The occurrence of 25 Losses of Capture between two
consecutive days
•
ERI: the device will be set to ERI mode with VCC OFF
•
Unipolar lead failure is detected
The occurrence of these unsuccessful, aborted or postponed
SA/CV sequences and disabling of the VCC feature are reported
in the Status log in the VCC statistics.
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......