![BIOTRONIK Evia DR Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/biotronik/evia-dr/evia-dr_technical-manual_2756148109.webp)
Evia Technical Manual 101
Table 26: Symptomatic First-Degree AV Block and
Infrequent Paroxysmal Pauses (< 5%)
Standard
Program
Symptomatic
First-Degree
AV Block
Infrequent
Paroxysmal
Pauses
(< 5%)
Valid for: DR(-T)
DR(-T)
SR(-T)
Parameters
Mode DDD-R
DDD
VVI
Rate Hysteresis
(ppm)
OFF OFF
-10
Rate Hysteresis
Rep. Cycles
--- ---
5
Rate Hysteresis
Scan Cycles
--- ---
5
AV Hysteresis
OFF
OFF
---
2:1 Lock-in
Protection
ON ON
---
Mode switch
ON
ON
---
PMT ON
ON
---
Holter settings
High atrial rate
AT
OFF
OFF
High ven. Rate
ON
OFF
OFF
Pre-trigger
recording (%)
75 ---
---
IEGM signal
Filtered
---
---
HAR limit (bpm)
200
---
---
HVR limit (bpm)
180
---
---
HVR counter
8
---
---
7.15 Home Monitoring (Evia DR-T)
Home Monitoring enables the exchange of information about a
patient’s cardiac status from the implant to the physician. Home
Monitoring can be used to provide the physician with advance
reports from the implant and can process them into graphical
and tabular format called a Cardio Report. This information
helps the physician optimize the therapy process, as it allows the
patient to be scheduled for additional clinical appointments
between regular follow-up visits if necessary.
The implant’s Home Monitoring function can be used for the
entire operational life of the implant (prior to ERI) or for shorter
periods, such as several weeks or months.
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......