154 Evia Technical Manual
15.2 Pulse- and Control Parameters
Basic rate
30...(1)...
60
...(1)...88...(2)...122...(3)...140...(5)...200 ppm
Night rate
Off
, 30...(1)...60...(1)...88...(2)...122...(3)...140...(5)...200 ppm
Rate Hysteresis
Off
; -5...(5)...-90 bpm
Repetitive Hysteresis
Off;
1…(1)…15
Scan Hysteresis
Off
: 1…(1)…15
Upper Rate
90…(10)…200 ppm
UTR Response
WRL (automatic selection)
Rate Limitation
4,5,6
190…220 ppm
Dynamic AV Delay (Dual chamber only)
Off;
low
; medium; high, (individual, or fixed), IOpt
4
The corresponding intervals t correlate with the rates f by the formula t = 60.000
/ f (t in ms, f in ppm).
5
In the event of electronic defect.
6
Rate Limitation changes as the Pacemaker approaches End of Service. The
Rate Limitation is nominally 190 ppm at Beginning of Service (BOS) and can
reach 220 ppm at End of Service (EOS) due to battery depletion.
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......