![BIOTRONIK Evia DR Скачать руководство пользователя страница 177](http://html1.mh-extra.com/html/biotronik/evia-dr/evia-dr_technical-manual_2756148177.webp)
Evia Technical Manual 169
The VDD mode is conditionally indicated for
•
patients with normal sinus rhythms and normal AV
conduction, but who intermittently need ventricular
pacing.
The VDD mode is
contraindicated
for
•
frequent or persistent supraventricular tachyarrhythmia,
including atrial fibrillation or flutter, and/or
•
inadequate intra-atrial complexes that do not permit safe
sensing, and/or
•
intact
retrograde
conduction.
Single Chamber Modes
VVI
The VVI mode is
clearly indicated
for
•
all symptomatic bradyarrhythmias, but particularly if
1. the atrium does not significantly contribute to the hemo-
dynamics (persistent or paroxysmal atrial flutter or
fibrillation, dilated atria).
4. there are no grounds for development of pacemaker
syndrome through loss of the atrial contribution or
through negative atrial contribution.
The VVI mode is conditionally indicated for
•
symptomatic bradycardia when the simplicity of the
pacing system is of crucial significance due to
1. senility (for the sole purpose of prolonging life).
2. incurable
illness.
3. great distance from the follow-up care center to the
patient's home.
4. absence of retrograde VA conduction.
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......