40 Evia Technical Manual
Table 7: Percentages of Patients Meeting Success Criteria
Criteria
% Patients
Slope > 0.650
53.8% (7/13)
Within 20 bpm of MCLR*
69.2% (9/13)
Achieved MCLR
46.2% (6/13)
* MCLR = Maximum Closed Loop Rate
Figure 4
shows the obtained heart rate versus the expected
heart rate during the CAEP treadmill test for all patients
completing at least 3 stages of exercise (n = 13).
0.00
0.20
0.40
0.60
0.80
1.00
1.20
0.00
0.22
0.39
0.54
0.73
1.00
CAEP Workload (Normalized)
Ob
tained
HR
(n
o
rm
a
lized)
Wilkoff Predicted HR
Lower 95% CI
Mean (n = 13)
Upper 95% CI
Figure 4. Obtained vs. Expected Heart Rate during CAEP