Evia Technical Manual 51
Figure 7: Median Time from Onset to Evaluation of
Arrhythmic Events
Analysis
The mean time from onset to evaluation of first AF, VT, and VF
events in Group 2 is greater than the mean time from onset to
evaluation of first AF, VT, or VF events in Group 1. A rejection
of the null hypothesis for AF, VT and VF event types indicates
that the mean time from onset to evaluation of the first AF, VT
and VF events in Group 1 is significantly less than the mean time
from onset to evaluation of the first AF, VT and VF events in
Group 2. P-values are =0.005, <0.001 and <0.001 respectively.
6.4.4 Conclusions
•
Use of HM in Group 1 resulted in an average of 1.9 office
visits per patient year in the 12 months post-implant, versus
an average of 3.4 office visits per patient year in Group 2, a
44% reduction in office visits. The average number of office
visits is significantly less in the HM group than in the Control
group (p < 0.001).
Содержание Evia DR
Страница 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Страница 8: ...vi Evia Technical Manual...
Страница 13: ...Evia Technical Manual 5...
Страница 15: ...Evia Technical Manual 7...
Страница 28: ...20 Evia Technical Manual...
Страница 128: ...120 Evia Technical Manual...
Страница 137: ...Evia Technical Manual 129...
Страница 143: ...Evia Technical Manual 135...
Страница 152: ...144 Evia Technical Manual...
Страница 156: ...148 Evia Technical Manual...
Страница 160: ...152 Evia Technical Manual...
Страница 167: ...Evia Technical Manual 159 Ongoing Atrial Episode 6 12 18 hours Event Report Off On Patient Report Off On...
Страница 182: ......