Perform the Tests
|
95
If any set of mean values is out of the expected range, review the
other three sets of mean values for the same channel pair. For
example, if the A1/H12 comparison fails (the wells are not within
the expected range of each other), review the comparisons of
A2/H11, H1/A12, and H2/A11. If two or more sets of mean values
for a channel pair are out of the expected range, there is a
problem with one of the eight read channels. If only one of the
four mean values results in a failure, check the well for debris and
the plate for scratches or fingerprints.
Accuracy Specification:
For comparison in this test, the following accuracy specification is applied,
using Normal reading mode and a 96-well microplate.
± 1.0% ± 0.010 OD from 0.000 OD to 2.500 OD at 405 nm
Perform Liquid Test 2
The recommended method of testing the instrument performance is to
use the Absorbance Test Plate to confirm alignment, repeatability, and
accuracy, which will also confirm linearity. If a Test Plate is not
available, Liquid Test 2 can be used for these tests.
Required Materials:
•
A new 96-well, flat-bottom microplate (Corning Costar #3590 is
recommended)
•
Ten test tubes, numbered consecutively, stored in a rack
•
Calibrated hand pipette (Class A volumetric pipette recommended)
•
Stock solution A or B (these are the same solutions as for Liquid Test 1)
•
A 0.05% solution of deionized water and Tween 20
Prepare Dilutions:
Create a percentage of dilution series, beginning with 100% of the original
concentrated stock solution (A or B) in the first tube, 90% of the original
solution in the second tube, 80% in the third tube, all the way to 10% in the
tenth tube. Dilute using the 0.05% solution of deionized water and Tween
®
20. This solution can also be made by diluting the BioTek wetting agent 200:1.
ELx808 Operator’s Manual
Содержание ELx808
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Страница 2: ......
Страница 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Страница 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Страница 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Страница 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Страница 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Страница 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Страница 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Страница 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Страница 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...