xiv
| Preface
BioTek Instruments, Inc.
Hazards
The following hazard warnings are provided to help avoid injury:
Warning! Internal Voltage
. Always turn off the power switch and unplug the
power cord or power supply before cleaning the outer surface of the instrument or
removing its top case.
Warning! Power Rating
. The instrument’s power supply or power cord must be
connected to a power receptacle that provides voltage and current within the
specified rating for the system. Use of an incompatible power receptacle may
produce electrical shock and fire hazards.
Warning! Electrical Grounding
. Never use a plug adapter to connect primary
power to the external power supply. Use of an adapter disconnects the utility
ground, creating a severe shock hazard. Always connect the power cord directly to
an appropriate receptacle with a functional ground.
Warning! Service.
Only qualified technical personnel should perform service
procedures on internal components.
Warning! Accessories.
Only accessories that meet the manufacturer’s
specifications shall be used with the instrument.
Warning! Liquids
. Avoid spilling liquids on the reader; fluid seepage into
internal components creates a potential for shock hazard or instrument damage.
Wipe up all spills immediately. Do not operate the instrument if internal
components have been exposed to fluid.
Warning! Unspecified Use
. Failure to operate this equipment according to the
guidelines and safeguards specified in this manual could result in a hazardous
condition.
Warning! Software Quality Control
. The operator must follow the
manufacturer’s assay package insert when modifying software parameters and
establishing reading methods.
Failure to conduct quality control checks could
result in erroneous test data
.
Warning! Reader Data Reduction Protocol
. When the reader is operated in
standalone mode the onboard assay software will flag properly defined controls
when they are out of range. The software will present the data with the
appropriate error flags for the operator to determine control and assay validity.
When operated via computer control, no limits are applied to the raw
measurement data. All information exported via computer control must be
thoroughly analyzed by the operator.
Warning! Hot Surface
. The lamp is hot when the instrument is turned on. Turn
off the reader and allow the lamp to cool for at least 15 minutes before attempting
to replace it.
Содержание ELx808
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Страница 2: ......
Страница 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Страница 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Страница 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Страница 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Страница 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Страница 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Страница 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Страница 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Страница 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...