Perform the Tests
|
91
Empty Carrier Test
Perform these steps (do
not
place a microplate on the carrier):
1.
For each filter in the filter wheel, run a simple endpoint protocol and then
print the results. (Do not apply blanking or transformations.)
2.
Examine the OD values on the printouts. Every OD should read 0.000
± 0.004 OD.
These limits are required by BioTek and cannot be changed.
3.
Sign and date the printouts, and store them with your test documentation.
If the Empty Carrier test fails at more than one wavelength:
•
Check the microplate carrier to ensure that it is clear of debris.
•
Verify that the lamp is properly aligned, and then run the test again. If the
lamp has been in use for approximately 600 hours, it may need to be
replaced, and the contacts in the lamp socket may need to be cleaned. Refer
to
Replace the Lamp and Clean the Contacts
in Chapter 5.
•
Follow the steps in
Inspect and Clean the Wavelength Filters
in
Chapter 5 to remove the filter wheel and clean the filters.
If the Empty Carrier test fails at just one wavelength:
•
Remove the filter wheel and examine the filter that exhibited the problem;
it may need to be replaced (contact BioTek). Check for spotting or a halo
effect.
ELx808 Operator’s Manual
Содержание ELx808
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Страница 2: ......
Страница 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Страница 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Страница 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Страница 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Страница 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Страница 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Страница 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Страница 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Страница 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...