| Chapter 5: Preventive Maintenance
Warning!
Hot surface. The lamp is hot when the instrument
is turned on. Turn off and unplug the reader. Allow the lamp
to cool down before attempting to replace it.
Important!
Oxide coating on the electrical contacts can
reduce lamp output. Clean the contacts in the socket
whenever the lamp is replaced (approximately every 500
hours).
You will need:
•
Tape
•
Screwdriver
•
Cotton swabs or pencil eraser
•
Isopropyl alcohol
•
Lamp, BioTek PN 3400508
To replace the lamp:
1.
Turn off and unplug the reader. Allow the lamp to cool for at least ten minutes.
2.
Temporarily tape the reader’s plate access door shut.
3.
Using your fingertips, locate the seven securing (perimeter) screws around the
sides and front where the reader’s top shroud meets the base. Remove the
screws using a screwdriver.
Tip:
Bring the reader to the edge of the work surface to access the
screws without having to turn the reader upside down.
4.
Carefully raise the front of the shroud and tip it toward the back of the
instrument.
5.
The lamp is located in the left rear quadrant inside the reader. Unplug the
lamp connector from the circuit board.
Important!
Do not touch the glass lens in the lamp
bracket.
6.
The lamp bracket is attached to the side of the instrument with two screws,
which are accessed from the outer left side of the instrument. Using a
screwdriver, loosen (do not remove) these two screws. Slide the bracket out of
the instrument.
BioTek Instruments, Inc.
Содержание ELx808
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Страница 2: ......
Страница 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Страница 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Страница 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Страница 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Страница 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Страница 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Страница 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Страница 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Страница 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...