background image

5

Prepare the System

Mount Column Holders

The system is shipped with column holders for Sfär columns 
of different sizes. Ensure that they are mounted in the desired 
positions on the system. 

Figure 11 shows how to mount a column holder on the right side 
of the system. There are also two positions for column holders 
on the front of the system that can be used for Biotage columns 
up to 50g; see Figure 12

Figure 11.

 Slide the column holder screw into the desired position on  

the right side of the system and then place the holder over the screw  

and fasten.

Figure 12.

 Biotage columns up to 50g can be placed on the front of  

the system. 

Inspect the Tubing

Warning

 

»

Shut down the system before replacing any tubing. Use only 

tubing designed for the Selekt Enkel system and supplied by 

Biotage.

Inspect all tubes before each run to ensure that they do not 
show signs of wear or damage and that they are properly 
connected and tightened. Use caution when finger tightening 
fittings to prevent stripped threads or crushed ferrules. 

All external tubing on the system except for the tubing on the 
collection arm can be replaced by the user. 

Note:

 All tube types, dimensions, and lengths are essential 

for the performance of the system. Only replace tubes with the 
equivalent tubes designed for the Selekt Enkel system and 
supplied by Biotage. 

Prepare the System

Start Up the System

Turn on the system using the power switch located on the left 
side of the system; see Figure 13.

Figure 13.

 The power switch is located on the left side of the system.

Assign Solvents and Set Reservoir Volumes

When a purification is run, the software determines which 
solvent inlets are connected to the solvents used in the run.

Note:

 For reversed-phase purification with methanol and  

water, it is strongly recommended to premix the water with  
5% of methanol and degass either through vacuum or 
sonication. Degassing of protic solvent blends decreases 
out-gassing of entrapped air during gradient elution, which  
will impact gradient performance and flow rates.

Note:

 With solvent and waste monitoring enabled (see 

page 24), the capacity and current fluid level must be entered 
each time you empty a waste reservoir or refill a solvent. 

Note: 

Both solvent inlets must be primed with solvent to 

achieve the specified pump performance.

1.  In the software, press 

Menu

 and then 

Solvent Setup

.

2.  Assign a solvent to each solvent inlet using the solvent 

drop-down lists (see Figure 14). 
To add solvents to the list, press 

Solvent Administration

.  

For more information, see page 22.

3.  If solvent monitoring is enabled, enter the capacity and  

the current solvent level of each solvent reservoir.

4.  If waste monitoring is enabled, enter the capacity and  

the current waste level of the waste reservoir.

5.  Prime the solvent inlets that have been assigned a new 

solvent; see below.

Содержание Selekt Enkel

Страница 1: ...Biotage Selekt Enkel User Manual...

Страница 2: ...nload the Run 15 Flush the System 15 Shut Down the System 16 Results 16 Search the Results 17 Analysis of the Result 18 Customize the Report 18 Add a Report Note 19 Export the Report 19 Base a New Run...

Страница 3: ...columns have a QR code that can be read by the QR reader underneath the touch screen see Figure 2 When scanning a column using the QR reader the system will be able to trace all runs performed on the...

Страница 4: ...n on screen notification when a solvent level is below 20 of the set capacity Issue an on screen notification when the waste level is above 85 of the set capacity Pause the system when it is time to r...

Страница 5: ...test tube or bottle reaches the specified vessel volume or a user defined maximum fraction volume for the run At any time you can manually switch to a new collection vessel by pressing in the chromat...

Страница 6: ...for collection and fractionation UV Baseline Correction When the UV baseline correction option is turned on the gradient run is preceded by a light absorbance detection phase During this phase the lig...

Страница 7: ...d supplied by Biotage Prepare the System Start Up the System Turn on the system using the power switch located on the left side of the system see Figure 13 Figure 13 The power switch is located on the...

Страница 8: ...purification and fill them with new solvents Note Both solvent inlets must be primed with solvent to achieve the specified pump performance Note We recommend to have both solvent reservoirs on the sa...

Страница 9: ...n the Rack panel or automatically identified by RFID when placed on the collection bedR See more in Specify the Rack Parameters on page 10 6 Set up the run parameters as described in the following sec...

Страница 10: ...nol and ethanol very volatile modifiers such as diethyl ether or modifiers that permanently alter the properties of the silica Reversed phase TLC is not supported Slight variations in migration rates...

Страница 11: ...see Figure 21 If the column has been scanned the column s ID and the number of times it has been used is also displayed For more information see Administrate Column Types on page 22 Figure 21 The Colu...

Страница 12: ...ne in the corresponding Select Rack drop down list or clear the check box in front of the rack Rack Sharing The system offers the possibility to have the first fraction s of a run collected in the las...

Страница 13: ...nt and waste monitoring is enabled you will be notified if there is not enough solvent or waste capacity for the run when you press 3 When the Sample Load dialog opens load your sample 4 Start the gra...

Страница 14: ...es a Spektra software license Baseline Flush After baseline detection the system is flushed with the solvent mix used at the start of the gradient Gradient The system is running a purification Line Fl...

Страница 15: ...will be resumed in the vessel with thicker border If all vessels are empty press Empty Rack reinserted after the run was resumed Empty vessels cannot be used If all vessels are empty press Empty The...

Страница 16: ...age 9 All changes except for changes to the gradient are instant When editing the gradient the modified gradient is purple and the ongoing gradient is black see Figure 32 The modified purple gradient...

Страница 17: ...lumn inlet and outlet tubes together Unload a Purification Flush the System If the last purification of the day is performed with a halogenated solvent e g DCM we recommend that you assign methanol or...

Страница 18: ...using the QR reader the system will be able to trace all runs performed using a specific column based on its ID To see all runs that have been performed using a column 1 Scan the column 2 Expand the...

Страница 19: ...olume handle is positioned is displayed above the handle If the gradient is shown the solvent mix is also displayed Select a collection vessel in the rack purple border and the volume handle in the ch...

Страница 20: ...mize a report by selecting it in the Results view and pressing Report Setup In the Report Setup view see Figure 42 the following content can be enabled or disabled in the report The changes are shown...

Страница 21: ...ct whether to export the full Selekt Enkel report or a customized version by turning the Apply Report Setup option off or on To customize the report see page 18 3 Press Export 4 In the Export dialog p...

Страница 22: ...ematic of the flow paths in Figure 45 2 Ensure that the correct column is chosen The column in the current setup will be chosen in the Column Type field If you want to change this select the column mo...

Страница 23: ...olume is the recommended volume for the selected column if one is selected 3 Press Flush 4 After the flush press Purge if needed to release any pressure remaining in the system Purge Use the purge fea...

Страница 24: ...ssure The safety pressure in bar or psi see page 24 If reached the flow rate will be reduced to keep the pressure at this level Equilibration Pressure The maximum pressure during the equilibration in...

Страница 25: ...ystem Owner and the password 1234 Before you log out it is strongly recommended that this password is changed User Parameters Visible If enabled the user account is available for use Name The user nam...

Страница 26: ...0 When the waste level is at 95 of the set capacity the system will be paused and the user will be prompted to empty the waste reservoir Figure 50 85 20 Level notification for waste and solvent Valley...

Страница 27: ...ing UV Zero Optional Detectors Biotage Solvent Detector Enable this option when the optional solvent detector s from Biotage is connected to the system Leak detection using up to three solvent detecto...

Страница 28: ...the screen turn off the system The power switch is located on the left side of the system 8 Turn on the system Note When the system is connected to your network the used IP address is displayed in the...

Страница 29: ...use a subfolder located in the share folder by adding it to the path e g server123 Biotage Selekt where server123 is the host name Biotage is the name of the share folder and Selekt is a subfolder In...

Страница 30: ...g Menu and then System Settings 3 Enable the E mail option in the Network field 4 Press the corresponding Settings button and enter the required e mail settings see Figure 54 Please contact your IT de...

Страница 31: ...onfirm Backup dialog opens Start the backup to the USB memory device by pressing Yes 6 When a message saying that the backup has been completed appears remove the USB memory device if applicable and p...

Страница 32: ...damage electronic components Note Avoid harsh cleaners and chemicals and moisture getting into the system 1 Shut down the system as described on page 15 2 Disconnect the power cord from the power out...

Страница 33: ...e the highlighted section in Figure 58 b Close the retaining nut Tip At the last couple of turns hold the flow cell flat against the retaining nut while closing the nut completely Note The retaining n...

Страница 34: ...nufacturer s safety handling storage and disposal recommendations refer to the safety data sheets SDS Electrical equipment can introduce ignition hazards Ensure that all solvent manufacturers recommen...

Страница 35: ...slide out the two brackets at the rear of the system see Figure 65 Reassemble by reversing the procedure Figure 64 B B A A Secondary solvent containment with a solvent detector A cable locking plate a...

Страница 36: ...holder Replace the Needle 1 Shut down the system as described on page 15 2 Remove the needle 3 Assemble the new needle ferrule and peek nut P N 413245SP Ensure that the needle is flush with the end o...

Страница 37: ...l not be registered as no light is reaching the detector at those particular wavelengths If using a system with a Spektra software license try performing the run with the UV Baseline Correction option...

Страница 38: ...n acceptable level before proceeding with the run using the lower flow rate Restart after power failure or shutdown with overpressure Warning When releasing a stuck check valve there is a risk of a sm...

Страница 39: ...eliver solvent The problem may be due to No or insufficient solvent in the reservoir s Refill the solvent reservoir s and ensure that both solvent inlet lines are submerged in solvent see Prime the So...

Страница 40: ...0 04 0 210 1 5 103 4 Dichloromethane DCM 75 09 2 0 42 5 V 235 6 9 475 3 Ethanol 64 17 5 0 88 2 II 210 1 3 90 0 Ethyl acetate 141 78 6 0 58 6 VIa 255 1 4 98 3 n Heptane 142 82 5 0 01 0 210 0 7 47 4 n...

Страница 41: ...t 1 412422 Biotage SNAP 750g 1500g Holder Assembly 1 412537 Male Luer Outlet Fitting for Biotage SNAP 750g and 1500g 1 412358 Biotage SNAP XL Fitting Female Luer Assembly 1 416331SP Flow Cell Selekt 1...

Страница 42: ...40 Biotage Selekt Enkel User Manual Biotage 2022 Notes Notes...

Страница 43: ......

Страница 44: ...art Number 417510 C 2022 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without noti...

Отзывы: