background image

26

Biotage

®

 Selekt Enkel User Manual | © Biotage 2022

Configure Network Share and E-mail Functionality

Configure Network Share and E-mail Functionality

Note:

 The Network Upgrade Kit (P/N 417501SP) is required for 

the network share and e-mail functionality.

Configure a Network Connection

N

To enable network share and/or e-mail functionality, start 
by setting up a network connection for your system as 
described below.

Note:

 Only users with system owner privilege can configure a 

network connection, network share, and e-mail functionality. 

1.  Connect a shielded category 5 TP cable between the 

ETH

 

port at the rear of the system and your network.

Note: 

Do not connect the network cable to the 

AUX

 port.

2.  Log in to the system settings view (see Figure 49 on 

page 24) by pressing 

Menu

 and then 

System Settings

.

3.  Press the 

Settings

 button next to the 

Network

 option in 

the 

Network

 field and enter the required network settings; 

see Figure 51. Please contact your IT department for help 
with entering the correct settings for your network.

Note: 

The 

Host Name 

will be used to identify the system on 

your network.

4.  When done, press 

Save.

5.  Enable the 

Network

 option in the 

Network

 field.

Figure 51.

 The Network Settings dialog using a static (left) or dynamic 

(right) setup.

6.  In the 

Restart Required 

dialog, press 

Shut Down

Note: 

The system has to be restarted for the new network 

settings to take effect.

7.  When the message saying that it is safe to turn off the 

system appears on the screen, turn off the system. 
The power switch is located on the left side of the system.

8.  Turn on the system.

Note: 

When the system is connected to your network, the used 

IP address is displayed in the 

About

 view.

Note:

 If you disable the 

Network

 option, the latest network 

settings are saved until enabled again.

Configure Network Share

N

With the system connected to the network, it is possible to give 
the system access to a network share folder. The share folder 
can be used for: 

 

»

Backup and restore of the system’s database.

 

»

Export of reports and system logs.

 

»

Automatic export of the report each time a run 
is completed, i.e. the reports are saved both on 
the system and in the network share folder.

Example on How to Create a Share Folder 

Using a Windows 10 Computer

Note: 

This procedure requires administrator permission on the 

Windows 10 computer.

1.  Create or select a folder that you want to share. The folder 

can be located anywhere on the computer.

2.  Right-click the folder and select 

Properties 

from the right-

click menu.

3.  In the 

Properties

 dialog, select the 

Sharing

 tab (see 

Figure 52).

4.  In the 

Sharing

 tab, click 

Advanced Sharing 

(see Figure 52)

5.  In the 

Advanced Sharing

 dialog, select the 

Share this folder 

check box and click 

Permissions 

(see Figure 52).

6.  Add the users that you want to give access to the folder and 

give them at least the permission to 

Change

 and 

Read

7.  When done, click 

OK

 twice to close the 

Permissions

 and 

Advanced Sharing

 dialogs.

8.  In the 

Properties

 dialog, make a note of the 

Network Path

This is the network path that you must enter when giving the 
system access to the share folder; see below.

9.  Click 

Close

 in the 

Properties 

dialog. 

Содержание Selekt Enkel

Страница 1: ...Biotage Selekt Enkel User Manual...

Страница 2: ...nload the Run 15 Flush the System 15 Shut Down the System 16 Results 16 Search the Results 17 Analysis of the Result 18 Customize the Report 18 Add a Report Note 19 Export the Report 19 Base a New Run...

Страница 3: ...columns have a QR code that can be read by the QR reader underneath the touch screen see Figure 2 When scanning a column using the QR reader the system will be able to trace all runs performed on the...

Страница 4: ...n on screen notification when a solvent level is below 20 of the set capacity Issue an on screen notification when the waste level is above 85 of the set capacity Pause the system when it is time to r...

Страница 5: ...test tube or bottle reaches the specified vessel volume or a user defined maximum fraction volume for the run At any time you can manually switch to a new collection vessel by pressing in the chromat...

Страница 6: ...for collection and fractionation UV Baseline Correction When the UV baseline correction option is turned on the gradient run is preceded by a light absorbance detection phase During this phase the lig...

Страница 7: ...d supplied by Biotage Prepare the System Start Up the System Turn on the system using the power switch located on the left side of the system see Figure 13 Figure 13 The power switch is located on the...

Страница 8: ...purification and fill them with new solvents Note Both solvent inlets must be primed with solvent to achieve the specified pump performance Note We recommend to have both solvent reservoirs on the sa...

Страница 9: ...n the Rack panel or automatically identified by RFID when placed on the collection bedR See more in Specify the Rack Parameters on page 10 6 Set up the run parameters as described in the following sec...

Страница 10: ...nol and ethanol very volatile modifiers such as diethyl ether or modifiers that permanently alter the properties of the silica Reversed phase TLC is not supported Slight variations in migration rates...

Страница 11: ...see Figure 21 If the column has been scanned the column s ID and the number of times it has been used is also displayed For more information see Administrate Column Types on page 22 Figure 21 The Colu...

Страница 12: ...ne in the corresponding Select Rack drop down list or clear the check box in front of the rack Rack Sharing The system offers the possibility to have the first fraction s of a run collected in the las...

Страница 13: ...nt and waste monitoring is enabled you will be notified if there is not enough solvent or waste capacity for the run when you press 3 When the Sample Load dialog opens load your sample 4 Start the gra...

Страница 14: ...es a Spektra software license Baseline Flush After baseline detection the system is flushed with the solvent mix used at the start of the gradient Gradient The system is running a purification Line Fl...

Страница 15: ...will be resumed in the vessel with thicker border If all vessels are empty press Empty Rack reinserted after the run was resumed Empty vessels cannot be used If all vessels are empty press Empty The...

Страница 16: ...age 9 All changes except for changes to the gradient are instant When editing the gradient the modified gradient is purple and the ongoing gradient is black see Figure 32 The modified purple gradient...

Страница 17: ...lumn inlet and outlet tubes together Unload a Purification Flush the System If the last purification of the day is performed with a halogenated solvent e g DCM we recommend that you assign methanol or...

Страница 18: ...using the QR reader the system will be able to trace all runs performed using a specific column based on its ID To see all runs that have been performed using a column 1 Scan the column 2 Expand the...

Страница 19: ...olume handle is positioned is displayed above the handle If the gradient is shown the solvent mix is also displayed Select a collection vessel in the rack purple border and the volume handle in the ch...

Страница 20: ...mize a report by selecting it in the Results view and pressing Report Setup In the Report Setup view see Figure 42 the following content can be enabled or disabled in the report The changes are shown...

Страница 21: ...ct whether to export the full Selekt Enkel report or a customized version by turning the Apply Report Setup option off or on To customize the report see page 18 3 Press Export 4 In the Export dialog p...

Страница 22: ...ematic of the flow paths in Figure 45 2 Ensure that the correct column is chosen The column in the current setup will be chosen in the Column Type field If you want to change this select the column mo...

Страница 23: ...olume is the recommended volume for the selected column if one is selected 3 Press Flush 4 After the flush press Purge if needed to release any pressure remaining in the system Purge Use the purge fea...

Страница 24: ...ssure The safety pressure in bar or psi see page 24 If reached the flow rate will be reduced to keep the pressure at this level Equilibration Pressure The maximum pressure during the equilibration in...

Страница 25: ...ystem Owner and the password 1234 Before you log out it is strongly recommended that this password is changed User Parameters Visible If enabled the user account is available for use Name The user nam...

Страница 26: ...0 When the waste level is at 95 of the set capacity the system will be paused and the user will be prompted to empty the waste reservoir Figure 50 85 20 Level notification for waste and solvent Valley...

Страница 27: ...ing UV Zero Optional Detectors Biotage Solvent Detector Enable this option when the optional solvent detector s from Biotage is connected to the system Leak detection using up to three solvent detecto...

Страница 28: ...the screen turn off the system The power switch is located on the left side of the system 8 Turn on the system Note When the system is connected to your network the used IP address is displayed in the...

Страница 29: ...use a subfolder located in the share folder by adding it to the path e g server123 Biotage Selekt where server123 is the host name Biotage is the name of the share folder and Selekt is a subfolder In...

Страница 30: ...g Menu and then System Settings 3 Enable the E mail option in the Network field 4 Press the corresponding Settings button and enter the required e mail settings see Figure 54 Please contact your IT de...

Страница 31: ...onfirm Backup dialog opens Start the backup to the USB memory device by pressing Yes 6 When a message saying that the backup has been completed appears remove the USB memory device if applicable and p...

Страница 32: ...damage electronic components Note Avoid harsh cleaners and chemicals and moisture getting into the system 1 Shut down the system as described on page 15 2 Disconnect the power cord from the power out...

Страница 33: ...e the highlighted section in Figure 58 b Close the retaining nut Tip At the last couple of turns hold the flow cell flat against the retaining nut while closing the nut completely Note The retaining n...

Страница 34: ...nufacturer s safety handling storage and disposal recommendations refer to the safety data sheets SDS Electrical equipment can introduce ignition hazards Ensure that all solvent manufacturers recommen...

Страница 35: ...slide out the two brackets at the rear of the system see Figure 65 Reassemble by reversing the procedure Figure 64 B B A A Secondary solvent containment with a solvent detector A cable locking plate a...

Страница 36: ...holder Replace the Needle 1 Shut down the system as described on page 15 2 Remove the needle 3 Assemble the new needle ferrule and peek nut P N 413245SP Ensure that the needle is flush with the end o...

Страница 37: ...l not be registered as no light is reaching the detector at those particular wavelengths If using a system with a Spektra software license try performing the run with the UV Baseline Correction option...

Страница 38: ...n acceptable level before proceeding with the run using the lower flow rate Restart after power failure or shutdown with overpressure Warning When releasing a stuck check valve there is a risk of a sm...

Страница 39: ...eliver solvent The problem may be due to No or insufficient solvent in the reservoir s Refill the solvent reservoir s and ensure that both solvent inlet lines are submerged in solvent see Prime the So...

Страница 40: ...0 04 0 210 1 5 103 4 Dichloromethane DCM 75 09 2 0 42 5 V 235 6 9 475 3 Ethanol 64 17 5 0 88 2 II 210 1 3 90 0 Ethyl acetate 141 78 6 0 58 6 VIa 255 1 4 98 3 n Heptane 142 82 5 0 01 0 210 0 7 47 4 n...

Страница 41: ...t 1 412422 Biotage SNAP 750g 1500g Holder Assembly 1 412537 Male Luer Outlet Fitting for Biotage SNAP 750g and 1500g 1 412358 Biotage SNAP XL Fitting Female Luer Assembly 1 416331SP Flow Cell Selekt 1...

Страница 42: ...40 Biotage Selekt Enkel User Manual Biotage 2022 Notes Notes...

Страница 43: ......

Страница 44: ...art Number 417510 C 2022 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without noti...

Отзывы: