Installation
Installation
Warning
»
Follow regional safety practices when handling and moving
shipping boxes and containers, and moving the system.
»
The system must be placed in a well-ventilated fume hood or an
equivalent enclosure that is capable of exhausting 2.5 m
3
/min
(90 ft
3
/min). Follow local and national safety regulations
for installing a system inside a fume hood and the safety
regulations supplied by the fume hood manufacturer.
»
Never install or use the system near a potential ignition source.
»
If the system is placed in a walk-in fume hood, a fume extractor
must be used for open waste/fraction collection vessels.
»
The total weight of the package including the system is
between 16 and 84 kg (35 and 185 lbs) depending on the
system configuration. Use suitable lifting equipment when
moving the package.
Note:
We recommend that the boxes and packing materials are
retained in case the system needs to be returned for service
or moved to another location. If you need to ship the system,
please contact Biotage 1-Point Support for instructions.
Set Up the Radial Compression Module
Warning
»
The outlet of the safety relief valve must be directed away from
the operator.
1. Remove the assembled radial compression module (see D in
Figure 13 on page 11) from the shipping container.
2. Attach the three-way valve supplied in the start-up kit onto
the top head assembly using a 9/16" wrench.
3. Determine the best installation location for the radial
compression module. Consider the following:
»
The outlet of the safety relief valve
(see Figure 3 on page 2) must be
directed away from the operator.
»
There must be sufficient space above the radial
compression module to insert and remove cartridges.
4. If using a Flash 75 radial compression module:
a. Screw the two mounting rods into the mounting pads
located on the side of the barrel.
b. Attach the barrel to the fume hood’s mounting bars
using mounting clamps (not supplied).
5. If using a Flash 150 radial compression module, roll the cart
with the module into a walk-in fume hood.
Set Up the AM-190 Manifold
Warning
»
The AM-190 manifold must be mounted horizontally so that the
water trap can function correctly.
»
The outlet of the safety relief valve must be directed away from
the operator.
1. Remove the AM-190 manifold (see A in Figure 13 on
page 11) from the shipping container.
2. Attach the two mounting bars to the manifold using a 5/32"
Allen key.
3. Determine the best mounting location for the manifold.
Consider the following:
»
Choose a location that allows you to easily access both
the manifold and the radial compression module.
»
The outlet of the safety relief valve (see B in Figure 4 on
page 3) must be directed away from the operator.
4. Attach the AM-190 manifold to the fume hood’s mounting
bars using mounting clamps. Ensure that the manifold is
mounted horizontally.
Set Up the Sample Injection
Module (Biotage
®
SIM)
Note that the use of the sample injection module (SIM) is optional.
1. Remove the SIM (see B in Figure 13 on page 11) from the
shipping container.
2. Open the bag of bottom frits and push one plastic frit into a
sealing ring. Save the other frits and sealing rings for use later.
3. Unscrew the lower half of the stainless steel frit holder
and drop the sealing ring and frit into the lower half; see
Figure 22 on page 16.
Note:
Ensure that the plastic frit is at the bottom of the
frit holder.
4. Screw the lower half of the frit holder back in place. Finger-
tight is normally adequate.
5. Determine the best mounting location for the SIM. Consider
the following:
»
Typically, the SIM should be mounted so that
the bottom tube fitting is 1" to 3" (25 to 76 mm)
above the radial compression module.
»
Ensure to leave enough space to insert and remove
cartridges from the radial compression module.
»
The SIM should be in a convenient location
for pouring in liquid or solid sample.
6. Attach the SIM to the fume hood’s mounting bars using the
chain clamp provided.