- 8 -
1.9 Electronic control
The whole electronic control system is contained in a rack system at the back right of the instrument. The
various functions are controlled by five electronic boards. Two of them control the mechanical movements.
The ADC board controls the photometer functioning and the signals conversion.
The temperature board controls the temperature and the functioning of level sensor.
The interface board communicates with the external management computer via serial cable.
The power supply of the instrument is located under the rack.
1.10 Management Computer System
The external computer managing the KENZA 240 is a standard computer, with at least 512 MB RAM, and a
hard disk of a minimum of 40 GB, a CD player and at least one COM1 serial port for connection to the
analyser. In the event of a further connection of the KENZA 240 to the laboratory management system –
LIMS – two serial ports (COM1 and COM2) are essential.
It is also necessary to have a CPU which supports Windows, 2000, XP, Vista/.7 operating system, given that
the software management runs with Windows. The Input/Output peripheral devices such as the keyboard,
mouse, display, speakers and printer will be connected to this computer. The external printer is important as
this instrument is not equipped with an internal printer.