Parameter
Description
Allowed value
Sample volume
Defines the sample volume (µL)
2-100 (step 1)
Water volume
Defines the water volume to be added to the
reaction mixture (µL)
0-300(step 1)
Sample washing
Defines the number of washes to be performed by
the sampling system after having sampled this test.
1-3
No. of Reag.
Defines the number of reagents used for this test.
1-2
MultiPoint
Defines if the calibration is performed with several
calibration points.
1 (Yes) or 0 (No)
STD number
Defines the standard position which shall be used to
calibrate the test. The standard positions can be
shown by clicking on the button “View STD
positions”.
1-8
Reag. 1 Vol.
Defines reagent 1 volume (µL)
2-350 (step 1)
Reag. 1 Pos.
Defines reagent 1 position. (Refer to “0
Quick selection of reagent position”, page - 25 -)
1-60
Vial R1.
Defines the type of container for reagent 1.
A: container of 40mL (medium, internal)
B: container of 17mL (small, external)
C: container of 60mL (large)
Reag. 2 Vol.
Defines reagent 2 volume.
2-350 (µL)
Reag. 2 Pos.
Defines reagent 2 position. (Refer to “0
Quick selection of reagent position”, page - 25 -)
1-60
Vial R2.
Defines the type of container for reagent 2.
A: container of 40mL (medium, internal)
B: container of 17mL (small, external)
C: container of 60mL (large)
STD Conc.
Concentration of standards.
Concentration. In case of multi-standard
calibration, the position 1 is reserved for
concentration 0.
Содержание KENZA 240 TX
Страница 5: ...5 1General Information...
Страница 10: ...10 2Installation...
Страница 14: ...14 3Preliminary operations...
Страница 17: ...17 4Methods programming...
Страница 30: ...30 5Calibrations and controls...
Страница 37: ...37 6Prepare the run...
Страница 47: ...47 7During the run...
Страница 51: ...51 8After the run...