5.4 View/Modify Controls
To prepare the control values (target range), click on
The following screen appears :
To program a control, select the control number, enter a Name and Lot number. Then enter the values (minimum
and maximum), and click on
.
5.5 Request controls
To optimize the use of positions on the sample tray, 2 kind
fixed positions. These controls are the most commonly used
C3 to C10 are used for specific controls: urine controls, HDL/LDL/CK
These positions are not fixed and have to be set.
This strategy has been chosen to avoid the waste of too many fixed positions.
N
N
o
o
t
t
e
e
: You can also program the common controls (usually programmed in C1 and C2) in C3 and C4 to carry
out the intra-serial controls. In this case, C1/C3 and C2/C4 can be linked to get the charts with all values.
5.5.1 C1 or C2
C1 and C2 –used for “classical” multiparametric controls
These controls are requested automatically with the calibration.
5.5.2 C3 to C10
To run controls C3 to C10, enter the patient menu and program
View/Modify Controls
To prepare the control values (target range), click on
, then
.
program a control, select the control number, enter a Name and Lot number. Then enter the values (minimum
.
To optimize the use of positions on the sample tray, 2 kind of controls have been developed : C1 and C2 are at
fixed positions. These controls are the most commonly used.
C3 to C10 are used for specific controls: urine controls, HDL/LDL/CK-MB controls, NH3/EtOH/CO2 controls, …
e to be set.
This strategy has been chosen to avoid the waste of too many fixed positions.
: You can also program the common controls (usually programmed in C1 and C2) in C3 and C4 to carry
serial controls. In this case, C1/C3 and C2/C4 can be linked to get the charts with all values.
C1 or C2
parametric controls – are fixed in positions C1 and C2 of the tray.
These controls are requested automatically with the calibration.
C3 to C10
To run controls C3 to C10, enter the patient menu and program a position as control as follows :
Select the control number.
The controls information is automatically filled (ID
and Name).
Then select the methods to be controlled.
- 34 -
program a control, select the control number, enter a Name and Lot number. Then enter the values (minimum
of controls have been developed : C1 and C2 are at
MB controls, NH3/EtOH/CO2 controls, …
: You can also program the common controls (usually programmed in C1 and C2) in C3 and C4 to carry
serial controls. In this case, C1/C3 and C2/C4 can be linked to get the charts with all values.
are fixed in positions C1 and C2 of the tray.
a position as control as follows :
Select the control number.
The controls information is automatically filled (ID
Then select the methods to be controlled.
Содержание KENZA 240 TX
Страница 5: ...5 1General Information...
Страница 10: ...10 2Installation...
Страница 14: ...14 3Preliminary operations...
Страница 17: ...17 4Methods programming...
Страница 30: ...30 5Calibrations and controls...
Страница 37: ...37 6Prepare the run...
Страница 47: ...47 7During the run...
Страница 51: ...51 8After the run...