18
Mixing Properties
Specifies whether the mixing is performed
Online
(by the instrument during a run) or
Offline
(manually).
Once
Type
is set to
Mix
, the
Mixing Components
link will be available in the top right corner of the
Reagent Details
box.
Use the
Mixing Components
screen to designate the volume (in μl) of chromogen buffer and the volume (in μl) of
chromogen substrate. The available components are listed on the left; the components to be mixed are listed on the
right. Total volume of mixing components cannot exceed 6ml.
1. From the
Chromogen Components
list on the left, highlight the catalog number for chromogen buffer
and click the
Add
button to add to the list on the right.
2. Click on the
Volume
field drop-down menu to select the desired volume (in μL) of buffer.
3. Add
Enzyme Components
to the list in the same way. The minimum volume which may be selected
is 100 μl.
4. To delete a component from the list on the right of the screen, highlight it in either list and click the
Subtract
button.
5. Click
Save
, then click
Close
to exit.
Counter Stain Group
The Counterstain reagent group includes all counterstains. This screen will allow the addition, modification,
and deletion of counterstains. At the bottom of the
Reagent Details
box the following information is displayed:
Waste Disposal Properties:
Specifies whether the used reagent is considered hazardous.
Note: The designation of Hazardous or Non Hazardous is dependent upon local and national government
regulations.