6
Environmental Conditions
Placement
Indoor
Operating temperature range
15-30°C (59-86°F). Recommended temperature for optimal staining is 68-75°F
Min. to Max. Temperature range
10-40°C (50-104°F)
Maximum relative humidity
80%, Non-condensing
Ventilation
Do not block the vent openings on the instrument. Leave a minimum of 5 cm (2 in.) of space between the automated
slide stainer and the wall, cabinets or other obstructions.
Venting (exhaust)
The instrument does not need special exhaust ventilation, unless required by safety regulations when using certain
reagents.
Humidity
Avoid humidity extremes. Low humidity can cause electrostatic problems that may affect the computer. High
humidity may create condensation.
Temperature
The intelliPATH automated slide stainer must be placed in a location away from direct sunlight and extreme heat
or cold.
Water Quality
The intelliPATH requires CLSI Type II water or equivalent (referred to as deionized water).
Support
To contact Biocare Medical Technical Support by telephone, please call 1-800-799-9499, Option 3.
You may also email Technical Support at [email protected] or submit a message online at www.biocare.net