2
Transportation
Shipping environment is 0-50ºC (32-122ºF). Do not stack. Protect from moisture. In addition to the ASTM regulation required markings,
label the shipping container with shipping and storage temperatures as well as a warning to not stack and protect from moisture.
Storage
Storage environment is 10-50ºC (50-122ºF), 10-80% non-condensing relative humidity. Do not stack. Protect from moisture. Not to
exceed one year (12 months) in storage.
Handling
The Decloaking Chamber NxGen may be safely handled by one person capable of lifting 20 pounds (9.1 kg). Lift by the handles only.
Do not move unless unit is completely cooled.
Operating Environment
Standard Laboratory Conditions: 15-30ºC; 15-80% relative humidity. Unit must be placed on a level bench-top capable of safely
supporting the weight of the Decloaking Chamber NxGen and contents. A minimum of 2” clearance on all sides is suggested for
adequate ventilation. High Altitude Recommendation: At greater than 3000M, increase time suggestions.
Power Requirements
The Decloaking Chamber NxGen must be plugged into a properly grounded power outlet. Do not defeat the grounding provisions by
bypassing the grounding pin or the power cord
.
If necessary the Decloaking Chamber NxGen may be plugged into a suitable, UL/CSA/
CE approved power strip providing it is the only device connected and the power strip has a minimum electrical rating greater than
1000 Watts.
USB Flash Drive
The Decloaking Chamber NxGen is equipped with a USB flash drive that will create and save a data file for every run. The USB flash
drive feature requires the use of a USB 2.0 flash drive with a minimum capacity of 512 MB and a maximum capacity of 4 GB.
Waste Disposal
All reagents used in the Decloaking Chamber NxGen must be properly disposed of in accordance with local, state, and federal
regulations.
Instrument Disposal
Dispose of the Decloaking Chamber NxGen in accordance with local, state, and federal regulations. Decontaminate with an antibacterial
cleansing agent.
Accuracy
Temperature readings are accurate /- 5ºC. Pressure readings are accurate /- 3%.
Warranty
The Decloaking Chamber NxGen warranty is for one year (12 months) from receipt of purchase and covers all parts and labor when
performed solely by Biocare Medical. Warranty is invalidated if equipment is abused, damaged, or improperly maintained by customer.
Warranty is not transferable to any other party should the equipment be resold or transferred by the customer to another party. To the
extent permitted by law, Biocare Medical disclaims any liability for any incidental or consequential damages related to this equipment
or for any warranty related services it performs.
Safety Precautions
Decloaking Chamber NxGen infectious sample use is limited to tissues/specimens that are specified on Biocare’s product datasheets.
Handle control slides and test FFPE slides as if capable of transmitting infectious agents, and dispose with proper precautions. It is
recommended that the specimens are handled using established good laboratory working practices. No flammable or corrosive solvents
are to be used in the Decloaking Chamber NxGen.
Service
Units under warranty requiring service will be returned to Biocare Medical for depot repair services. No field repairs or firmware updates
will be performed on this unit. Except for the sealing gasket kit, there are no user serviceable parts.
Technical Support: 1-800-799-9499 Option 3