15
Quick Instruction Guides
Instrument Operation
1. Plug in the Decloaking Chamber NxGen and turn ON / OFF switch to the ON position (|).
2. Add 500ml of deionized or filtered water and a metal rack holder to the water chamber.
Caution: Less than 500ml of water may cause adverse staining results.
3. Add 3 metal slide canisters containing slides and antigen retrieval solution (or deionized / filtered water) to the metal rack holder.
4. Lock the lid into place. Ensure that pressure cable is attached.
5. Select program temperature.
6. Select program time.
7. Insert a USB drive if data recording is desired.
8. Push the green button at pre-run screen to start the protocol.
9. Upon run completion and unit is under 95ºC, carefully open the lid. Tilt the lid at 45º and gently shake to drain
excess water before fully removing the lid. Use extreme caution as steam and liquids can still be very hot!
10. Remove metal rack holder by grasping the plastic handles. Do not grab the metal rack holder or the metal slide canisters by
the metal – it is very hot!
11. Continue with slide processing.
12. Once the unit is cool, remove water chamber and properly dispose of water. Dry outside completely before placing in unit.
13. Leave lid slightly ajar or next to Decloaking Chamber NxGen.
Software Reports
1. Transfer USB key from Decloaking Chamber NxGen to PC.
2. Open Decloaking Chamber NxGen Reports Software.
3. Import files from USB flash drive to PC.
4. Select desired reports and appropriate parameters for report format.
5. Print or export reports for record keeping purposes.