95°C Protocol Function
The Decloaking Chamber NxGen has the ability to preheat retrieval solutions when the 95°C protocol is selected. The preheat functionality
temporarily pauses the unit when the target temperature is reached, allowing the user to add deparaffinized slides to retrieval buffers
that have already been warmed to 95°C. This workflow modification enables better compatibility with protocols for breast tissue retrieval
that previously required the use of a water bath for preheating solutions.
The unit will beep when the target temperature has been achieved. The following message will be displayed (pictured below) on the unit
screen with two separate options for user input.
1. If slides will be added to the hot retrieval solutions, press the button under the “Load slides” option and open the unit to add the
necessary slides. Replace and secure lid. The run will resume automatically with the user selected hold time after the unit temperature
equilibrates back to 95°C.
2. If preheat is not needed and slides are already present, nothing needs to be done. After 2 minutes the program will automatically
proceed to the run screen and deduct the two minutes from the programmed hold time. If desired, press "Skip" to immediately move to
run screen. Any elapsed time will deduct from the programmed hold time.