13
Maintenance
Daily
1. Make sure the power cord is unplugged before performing maintenance.
2. Pour the water out of the metal water chamber, rinse with deionized or filtered water and dry. Do not use harsh chemicals or scouring pads.
Wipe down the outside of the metal water chamber, including the bottom. Examine the metal water chamber rim and bottom
for dents or deformations. Call Biocare Technical Support to report any concerns.
3. Remove the condensation collector and empty of any accumulated water.
4. Leave lid next to unit or slightly ajar atop unit.
5. Do not leave liquids in the Decloaking Chamber NxGen when not in use.
6. Wash the metal slide canisters in soap and water. Do not use bleach.
Monthly
1. Make sure the power cord is unplugged before performing maintenance.
2. Unplug the pressure cable from the lid.
3. Clean the outer body with a soft cloth (paper towel or microfiber cloth). Do not immerse outer body in water or pour water into it.
4. With warm water rinse the underside of the lid including the sealing ring, pressure limit valve, anti-block cover, air escape and
pressure indicator poppit. Dry completely.
5. Clean area under upper ring with dampened cloth or microfiber cloth. Do not use chemical cleaners.
6. To clean sealing ring, hold the knob on the sealing ring supporting cover and pull the sealing ring up. After cleaning, put the sealing
ring supporting cover back. Note: The side with the knob should face outward. Do not put it upside down (Figures a and b).
a
b
7. Inspect the rubber grommet for wear. Carefully remove the sealing ring supporting cover and inspect the grommet. Replace with
the sealing gasket kit if worn.
8. To clean rubber grommet located in center of sealing ring supporting cover, carefully remove the sealing ring supporting cover
and clean the grommet; then reinstall it.
9. To clean the pressure limit valve, remove and rinse it with water, then reinstall it.
10. To clean the condensation collector, remove it and clean it with warm soapy water, then reinstall it.
11. Reconnect the pressure cable to the lid.