Beurer GL50 evo

4

Remove and dispose of lancet needle 

Press the trigger again so that you can grip the shaft. Carefully remove 

the lancet needle from the lancing device and dispose of the lancet in 

a puncture-proof container.

Carefully dispose of all blood samples and materials that came into 

contact with you or your patients. This prevents injuries and the 

infection of others.

5

Replace cap

Place the cap back on the device. 

5.6 Evaluating measured blood glucose values  

Your blood glucose monitor can process measured values between 20 and 630 mg/dL (1.1 and 

35.0 mmol/L). A warning message is displayed in the event of measured values lower than 20 mg/dL 

(1.1 mmol/L) and measured values higher than 630 mg/dL (35.0 mmol/L).

 Warning

•  If you suspect that the blood glucose results are incorrect, first repeat the test and, if necessary, 

perform a functional test using control solution. Seek medical advice if dubious results persist.

•  Seek medical attention immediately if your symptoms do not correspond to your measured blood 

glucose values and you have followed all instructions for the Beurer GL50 evo blood glucose monitor.

•  Do not ignore symptoms of too high/low blood glucose levels. Always seek medical attention!

Blood glucose values

The following tables list blood glucose values based on the STANDARDS OF MEDICAL CARE IN 

DIABETES 2016 from the ADA (American Diabetes Association).

Time of the blood glucose 

measurement

Normal blood 

glucose values

Increased risk of 

diabetes (prediabetes)*

Diabetes

On an empty stomach 

(fasting plasma glucose)

Below 100 mg/dL

Below 5.6 mmol/L

100 – 125  mg/dL

5.6 – 6.9 mmol/L

≥ 126 mg/dL

≥ 7.0 mmol/L

Two hours after an oral 

glucose tolerance test 

(consumption of 75 g)

Below 140 mg/dL

Below 7.8 mmol/L

140 – 199  mg/dL

7.8 – 11.0 mmol/L

≥ 200 mg/dL

≥ 11.1 mmol/L

* The risk increases continually, beginning with values below the lower limit of the range and increasing disproportionately towards 
the upper limit of the range.

Содержание GL 50 evo

Страница 1: ...3 IN 1 BLOOD GLUCOSE MONITOR Step by step GL 50 evo EN...

Страница 2: ......

Страница 3: ...ing a sample 21 5 3 Taking a blood sample and measuring the glucose level 22 5 4 Reading the result and marking measurements 24 5 5 Post processing and disposal 25 5 6 Evaluating measured blood glucos...

Страница 4: ...for self testing or in a clinical environment by trained personnel It enables you to measure your blood glucose quickly and easily store the measured values and display the average of all measured val...

Страница 5: ...the specified Customer Service address A B E D C A 3 in 1 monitor lancing device plug in USB and monitor B USB cover C 10 test strips D 10 sterile lancet needles 5 x 28G blue 5 x 33G purple E 1 practi...

Страница 6: ...t the measured values acoustically Display the average of the measured blood glucose values from the last 7 14 30 and 90 days Display the average of the marked measured blood glucose values from the l...

Страница 7: ...fficient for n tests LOT Batch designation Art Nr Order number Sterilised by radiation lancets Unit of measurement for blood glucose value Warning see accompanying documents Biohazard risk of infectio...

Страница 8: ...self testing Do not share the lancing device or lancet needles with others or amongst various patients risk of infection Use a new sterile lancet needle for each blood sample for single use only Gener...

Страница 9: ...ults Relative humidity of more than 90 may lead to inaccurate results Note The Beurer GL50 evo measuring system is intended for measuring capillary and venous whole blood Storage and maintenance Warni...

Страница 10: ...aste Electrical and Electronic Equipment If you have any questions please contact the local authorities responsible for waste disposal Notes on handling batteries If your skin or eyes come into contac...

Страница 11: ...ce 2 Plug in USB 3 Monitor Front Side 1 Display 2 Slot for test strips 1 2 1 Scroll wheel 2 ON OFF switch 3 Tensioning slider 1 2 3 3 2 Lancing device and lancet needles 1 Cap 2 Protective lancet disc...

Страница 12: ...ry 1 State of charge 2 Memory space number 3 Blood sugar unit mg dL or mmol L 4 Time 5 Symbol for marking measurements 6 Date 7 Measured value display Note The monitor is supplied with the following b...

Страница 13: ...information is followed Warning Use each test strip only once and for one patient only Handling test strips Note Securely close the test strip box immediately after taking out a test strip Do not use...

Страница 14: ...d film packaging never use other containers 4 Initial use and basic settings In order to activate the device move the ON OFF switch into the ON position on off The ON OFF switch is used as a keylock f...

Страница 15: ...eel until the arrow is pointing towards Settings then press the scroll wheel All values Average Settings Back 3 Rotate the scroll wheel until the arrow is pointing towards Date time then press the scr...

Страница 16: ...to set the timer minute 6 In order to start the timer rotate the scroll wheel to Start and press the scroll wheel Timer Start Stop 3 Setting the alarm You can set an alarm on the blood glucose monito...

Страница 17: ...iefly pressing and holding the scroll wheel 2 Rotate the scroll wheel until the arrow is pointing towards Settings then press the scroll wheel All values Average Settings Back 3 Rotate the scroll whee...

Страница 18: ...essing the scroll wheel Beep on Beep off Result beep Back Note The result beep transmits the measurement as an acoustic beep signal Example measured value 103 Start signal Beep beep Number 1 Short bee...

Страница 19: ...ng towards Turn display then press the scroll wheel Beep on off Ketone on off Turn display Language 8 Selecting the menu language The following menu languages are available English German French Spani...

Страница 20: ...fingertip To make the procedure as painless as possible do not take samples directly from the centre of the fingertip but slightly to either side Warning In the event of suspected hypoglycaemia take...

Страница 21: ...ve lancet disc by bringing it into a horizontal position and taking it off Retain the protective disc for the safe disposal of the used lancet needle after taking a blood sample 4 Place the cap onto t...

Страница 22: ...fore inserting it in the monitor 1 Prepare the test strip Take a test strip from the box film packaging and immediately close it again Use the test strip within three minutes of removal 2 Insert test...

Страница 23: ...from an alternative area Carry out another test with blood from your fingertip in order to confirm the test result If you often fail to notice that you have a low blood glucose level carry out a test...

Страница 24: ...e performs the blood glucose measurement The monitor counts down for approx five seconds The measurement is then shown on the display MEMORY MEMORY Read your measured value Check again that you have r...

Страница 25: ...ssing the scroll wheel 3 The selected marking is now allocated to the measured value and saved to the memory 5 5 Post processing and disposal 1 Remove test strip Remove the test strip from the device...

Страница 26: ...trol solution Seek medical advice if dubious results persist Seek medical attention immediately if your symptoms do not correspond to your measured blood glucose values and you have followed all instr...

Страница 27: ...spite the preprandial glucose targets having been reached Postprandial blood glucose measurements should be taken one to two hours after the start of a meal as this is when diabetics values are genera...

Страница 28: ...ely For the control solution test please observe the additional notes in the instructions for use for the control solution Important Do not use third party control solution Correct functioning of your...

Страница 29: ...te Never apply the control solution straight from the bottle to a test strip Reason the remaining solution in the bottle will be contaminated if the top of the bottle comes into contact with the test...

Страница 30: ...ible causes include Test strips were exposed to fresh air for too long The test strip box was not closed completely Repeat the test using a new test strip and or a new box of correctly stored test str...

Страница 31: ...s 4 You can cancel the process at any time To do so press and hold the scroll wheel or wait until the device switches itself off automatically after 2 minutes 6 2 Displaying average blood glucose valu...

Страница 32: ...on page 15 Measurements cannot be taken while the USB stick is connected to a PC The measurements remain saved on the blood glucose monitor when the USB stick is removed from the PC It is not possibl...

Страница 33: ...e device on different persons Do not submerge the device in disinfection solutions or any other fluids and make sure that no fluids can enter the device The cap on the integrated lancing device can be...

Страница 34: ...hout restriction Low temper ature Temperature of the measuring environment blood glucose monitor or test strip below the permitted range Repeat the test using a new test strip as soon as the measuring...

Страница 35: ...rvices Problem the test does not start after inserting the test strip into the device and applying blood Cause Solution Insufficient blood or test strip not filled correctly Repeat test using a new te...

Страница 36: ...ble for self testing System function test Each time device is switched on The serial number is located on the device or in the battery compartment EMC This device complies with the European standard E...

Страница 37: ...tead of blood The measurement using control solution must be within the result range This range is printed on each test strip box Chemical composition of the control solution The control solution is a...

Страница 38: ...1 1 4 2 135 6 7 5 136 3 7 6 1 4 0 1 1 1 3 351 5 19 5 350 8 19 5 2 8 0 2 0 8 System accuracy The GL50 evo blood glucose monitor in comparison to the YSI Three lots of GL50 evo blood glucose test strip...

Страница 39: ...within 15 of the values obtained in the medical laboratory with glucose concentrations of at least 100 mg dL 5 55 mmol L You will find further details and information regarding blood glucose results a...

Страница 40: ...opamine 2 2 mg dL 0 14 mmol L 5 0 mg dL 0 28 mmol L 1 0 EDTA 5 0 mg dL 0 17 mmol L 2 0 mg dL 0 11 mmol L 2 4 Ephedrine 40 mg dL 2 42 mmol L 3 9 mg dL 0 22 mmol L 2 4 Galactose 20 mg dL 1 11 mmol L 3 1...

Страница 41: ...l L 0 8 Tolazamide 2 5 mg dL 0 08 mmol L 2 3 mg dL 0 13 mmol L 1 8 Triglyceride 800 mg dL 9 37 mmol L 7 50 mg dL 0 42 mmol L 4 0 Uric acid 16 5 mg dL 0 98 mmol L 6 6 mg dL 0 37 mmol L 1 8 Xylose 9 5 m...

Страница 42: ...y claim shall only be considered if the buyer can provide Beurer or an authorised Beurer partner with a copy of the invoice purchase receipt and the original product The following are explicitly exclu...

Страница 43: ......

Страница 44: ...ncetas lancette SteriLance Medical Suzhou Inc No 168 PuTuoShan Road New District Suzhou 215153 China GL 50 evo Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer com www beurer healthguide...

Отзывы: