34
Beurer GL50 evo
8 w
hat
if
there
are
probleMs
?
Display messages on batteries and blood glucose measurement
No.
Cause
Solution
Empty
battery
Batteries discharged.
Charge the blood glucose monitor via the USB
interface.
High
temper-
ature
Temperature of the measuring
environment, blood glucose monitor
or test strip above the permitted
range.
Repeat the test using a new test strip as soon
as the measuring environment, monitor and
test strip have reached room temperature
(+20°C to +26°C). The check performed
at room temperature is used as a general
functional check. The operating range specified
in the technical specifications is valid without
restriction.
Low
temper-
ature
Temperature of the measuring
environment, blood glucose monitor
or test strip below the permitted
range.
Repeat the test using a new test strip as soon
as the measuring environment, monitor and
test strip have reached room temperature
(+20°C to +26°C). The check performed
at room temperature is used as a general
functional check. The operating range specified
in the technical specifications is valid without
restriction.
Strip
error
Used or contaminated test strip
inserted.
Insert an unused test strip that has not expired.
Repeat the blood glucose measurement.
Error 1
System error.
Contact Customer Services.
Error 2
Insufficient amount of blood on the
test strip.
Repeat the measurement using a new test
strip.
Error 5
System error.
Contact Customer Services.
Unknown error messages.
Contact Customer Services.