P
ROFESSIONAL
PROCEDURE
GUIDE
MULTI-TEST A1C SYSTEM
Preparation
1
2
3
4
5
7
9
8
6
BEFORE YOU BEGIN
COLLECT BLOOD
INSERT BLOOD
COLLECTOR
SHAKE
INSERT CARTRIDGE
DISPENSE SAMPLE
INTO CARTRIDGE
REUSE MONITOR
5 MINUTES TO RESULTS
PREPARE SAMPLER
Blood Collection
Blood Dilution
Blood Testing
Results
°C
°F
•
Run the test with all
parts of the test kit at
the same temperature
within the specified
range.
•
If the kit has recently been at high
temperatures (above 82ºF) or in the
refrigerator, keep the kit at room
temperature for at least one hour before
use.
•
Avoid running the test in direct sunlight,
on hot or cold surfaces, or near sources of
heat or cold.
•
Quality control materials should be used
to confirm the test kit is working properly.
Refer to the product insert for information
on when to run controls.
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
LOT 0628216
EXP 2014 01 27
0628216
Lot Number and Dating Label
Monitor (back)
Pouched
Test Cartridge
Ensure Lot #’s match
Complete
test within
15 minutes.
Sampler
Body
Blood
Collector
Base
(do not remove!)
WAIT T
O OPE
N
Or
45º
x
x
x
MATCH
CODES
SAMPLE
SAMPLE
SAMPLE
A1
A1
% A1C
A09
LOT 0628216
EXP 21 SEP 14
LOT 0628216
EXP 21 SEP 14
LOT 0628216
EXP 21 SEP 14
MA
TCH
CODES
% A1C
% A1C
MA
TCH
CODES
SAMPLE
SAMPLE
SAMPLE
A09
A09
% A1C
% A1C
SAMPLE
SAMPLE
SAMPLE
A09
% A1C
SAMPLE
SAMPLE
SAMPLE
A09
A09
% A1C
MATCH
CODES
Venous Draw
Mix blood
well before
testing
Too little
add more
blood
Not fully inserted
Keep pushing!
* If not, call Customer Service at
1-877-870-5610.
If you cannot resolve an error, please
call Customer Service at
1-877-870-5610.
Do not handle
Monitor again until
test is complete!
Do Codes
Match? *
Monitor and Test Cartridge codes
must match
Ensure Monitor is on level surface
WAIT
for SMPL to display
Ready for Sampler
Fully inserted
Just right
Too much
wipe away
excess
Collect blood from a slide
Push firmly!
Fingerstick
Shake well 6-8 times
This will mix the blood
with the solution
Push down completely to
dispense diluted sample
Remove quickly
Stand Sampler on table
while preparing Cartridge
“Click” Test Cartridge into place
Use your own
lancet device
to draw blood
Fully insert
Blood Collector into
Sampler Body
Keep
pushing!
Twisting
motion
helps
Gently touch blood
drop to fill
OPEN NOW
Use within 2 minutes.
Remove
Base
Display
counts down
QC test result*
# of tests left
Result
This result cycle remains displayed for
60 minutes or until the next Test Cartridge
is inserted.
THE MONITOR IS REUSABLE
To run another test, use a new
Sampler and Test Cartridge from the
same kit and return to Step 1,
“
PREPARATION.
”
ALWAYS MATCH LOT NUMBERS
Use Monitor only with the materials
included in the original kit. The Monitor
will expire after the programmed number
of tests have been run. If another Test
Cartridge is inserted, the Monitor will
display “00 TL.”
* If “QCOK” is not displayed, please see
list of error codes on reverse side.
Discard
Test Cartridge
Save
Monitor
Lot
Code
Product Dating
& Lot ID Label
Lot
Number
91078 B 3/2014