HemoCue Hb 201+: Quick Reference Guide

Hemocue Hb 201+ 

Quality Control (QC) is performed daily with patient use.  Monthly when not in use.

QC is completed within previous 24 hours then: 

Proceed to Patient Testing Steps 1-12 & 15-19

Hemocue Hb
Microcuvettes

EuroTrol HemoTrol 
QC Levels 1,2,3

QC is not complete then:

Proceed to QC Testing Steps 1-5 & 13-19.

QC solutions Level 1(Red); Level 2 (White); and Level 3 (Blue).

QC (3 levels) are within range before proceeding to patient testing.

Document QC results on the Hemocue Hb 201+ Record

Note:  Hemocue Hb Microcuvettes in singles is preferred at C&W.

Hemocue Hb 201+ Patient Test

Obtain Hemocue Hb 201+ Analyzer, Microcuvettes & Record Form.

Check that Hemocue Hb 201+ Microcuvettes are the Lot # in use & within expiry date.

Patient & QC Testing - Ready Hemocue Hb 201+   # 1-5

Note:  For Patient Testing, prepare analyzer before proceeding with Steps 1-5.

2

3

5

Use within 5 minutes

Use AC Adaptor or 
Batteries – 4 X type AA

Turn ON.  Previous Result 
is briefly displayed before 
“Hb” symbol.

Prepare analyzer to load

Pull out microcuvette tray

In 10 seconds display 
shows 3 flashing dashes.  

Self Test is passed.

Obtain Microcuvette from 
vial or indiviual package.

Close vial lid immediately.

Patient Test – # 6-12, # 15-19

Safe-T-Lance – Purple

Safe-T-Lance – Blue

Safe-T-Lance – PINK

6

7

Infant Heel 

1.6 mm depth x 1.5 mm 
width

Pediatric Finger

> 6 months & > 5 kg           
1.4 mm depth X 25 gauge

Finger > 8 yr – adult.       

1.2 mm depth x 1.5 mm width

Patient hand is warm.  
Use middle three fingers.  
Clean with disinfectant

Allow to air dry.          
Apply slight pressure on 
fingertip.

8

9

10

11

12

Choose puncture 
site. Puncture on 
side of fingertip 
and avoid nail 
bed.

Obtain puncture device as 
per age of patient.  Make 
the puncture.  Discard in 
sharps biohazard waste.

Wipe away the first two to 
three drops of blood with 2X2 
gauze.

Re-apply light pressure to 
the fingertip until another 
drop of blood is formed.

Touch open end of the 
microcuvette to drop of 
blood.  Fill in one 
continuous process.10 uL

Quality Control (QC) Test  (Levels 1, 2 & 3 are within QC range before patient testing.)

Dispense first drop onto a 
tissue.  Dispense second 
drop onto parafilm square.

Prepare QC solution:

Obtain 3 Levels QC solution.

Obtain separate parafilm square for each QC.

QC is at room temperature before use.

Check Lot # & Expiry Date; is current & in use.

Mix well just before use.

Touch open end of a 
microcuvette to QC drop. Fill 
in one continuous process.

13 

14

DO NOT REFILL 
microcuvette