c
4 Start-Up
30
Rev. 04, 03/2015
b
Configuring the Current Output on the Basic Module
(Continued)
5.
Set the desired position at the changeover switch (Sink Mode
(Fig. 12, Pos.1) or Position Source Mode (Fig. 12, Pos. 2).
1
Basic Module, changeover switch – Position Sink Mode
2
Basic Module, changeover switch – Position Source Mode (factory setting)
Fig. 12
Switching on the Basic Module Current Output at the Changeover
Switch
6.
Connect the module with the flat ribbon cable again.
7.
Push all the modules (Fig. 10, Pos. 3) of the measurement channel in again at
the same time.
8.
Fold the front panel up again.
9.
Tighten the screws again.
Note
Ensure that the position at the changeover switch on the basic mod-
ule and on the extension module is the same!
Use the same current output configuration on the basic and ex-
tension modules.
Содержание castxpert LB 452
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 14: ...c 2 Safety 14 Rev 04 03 2015 b ...
Страница 22: ......
Страница 40: ......
Страница 120: ......
Страница 138: ......
Страница 140: ...c 10 Decommissioning 140 Rev 04 03 2015 b ...
Страница 146: ...c 11 Technical Information 146 Rev 04 03 2015 b 11 4 c LB 452 Connections Rear Overview ...
Страница 147: ...c 11 Technical Information b Rev 04 03 2015 147 11 5 c LB 452 Connections Measuring channel base module ...
Страница 148: ...c 11 Technical Information 148 Rev 04 03 2015 b 11 6 c LB 452 Connections Measurement ChannelEx tension Module ...
Страница 151: ...c 12 Declaration of Conformity b Rev 04 03 2015 151 12Declaration of Conformity ...
Страница 152: ...c 12 Declaration of Conformity 152 Rev 04 03 2015 b ...
Страница 155: ...c 13 Index b Rev 04 03 2015 155 ...