c
4 Start-Up
26
Rev. 04, 03/2015
b
4.2
The Use of a Terminal Box
Install the terminal box in close proximity to the mould in a protected but easily
accessible place.
The GAMMAcast detectors and the EVU are to be connected via the terminal box.
This connection is to be made with a standard 6–wire shielded cable (6 x 0.5 mm²)
with a maximum length of up to 1000 m.
As required, the terminal box can be ordered with a plug-in connector (Id.
no. 34787, Fig. 8) or for open ends (Id. no. 07005, Fig. 9).
Fig. 8
Terminal Box for GAMMAcast Detectors with Plug-In Connector
(Id. no. 34787)
Fig. 9
Terminal Box for GAMMAcast Detectors with Open Ends
(Id. no. 07005)
Содержание castxpert LB 452
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 14: ...c 2 Safety 14 Rev 04 03 2015 b ...
Страница 22: ......
Страница 40: ......
Страница 120: ......
Страница 138: ......
Страница 140: ...c 10 Decommissioning 140 Rev 04 03 2015 b ...
Страница 146: ...c 11 Technical Information 146 Rev 04 03 2015 b 11 4 c LB 452 Connections Rear Overview ...
Страница 147: ...c 11 Technical Information b Rev 04 03 2015 147 11 5 c LB 452 Connections Measuring channel base module ...
Страница 148: ...c 11 Technical Information 148 Rev 04 03 2015 b 11 6 c LB 452 Connections Measurement ChannelEx tension Module ...
Страница 151: ...c 12 Declaration of Conformity b Rev 04 03 2015 151 12Declaration of Conformity ...
Страница 152: ...c 12 Declaration of Conformity 152 Rev 04 03 2015 b ...
Страница 155: ...c 13 Index b Rev 04 03 2015 155 ...