7 "Channel Overview" Menu
104
Rev. 04, 03/2015
b
7.3.1.2
Calibration of the Current Outputs
1.
Before calibrating the current outputs, connect an ammeter (not included in
the delivery contents) to the relevant current output.
2.
Click on
<Calibration>
(Fig. 65, Pos. 7).
A confirmation message opens.
3.
Click on
<Next>
in the confirmation message.
The device switches to test mode.
A command prompt opens.
4.
Click on the input field to enter a value for the 4 mA calibration point.
The numeric keypad opens (Fig. 18).
5.
Enter the value transmitted by the ammeter.
6.
Click on
<Next>
in the confirmation message.
7.
Carry out the calibration for 20 mA calibration point in the same way.
A confirmation message with the notification "Calibration finished"
opens.
8.
Click on
<Next>
to finish the calibration.
Содержание castxpert LB 452
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 14: ...c 2 Safety 14 Rev 04 03 2015 b ...
Страница 22: ......
Страница 40: ......
Страница 120: ......
Страница 138: ......
Страница 140: ...c 10 Decommissioning 140 Rev 04 03 2015 b ...
Страница 146: ...c 11 Technical Information 146 Rev 04 03 2015 b 11 4 c LB 452 Connections Rear Overview ...
Страница 147: ...c 11 Technical Information b Rev 04 03 2015 147 11 5 c LB 452 Connections Measuring channel base module ...
Страница 148: ...c 11 Technical Information 148 Rev 04 03 2015 b 11 6 c LB 452 Connections Measurement ChannelEx tension Module ...
Страница 151: ...c 12 Declaration of Conformity b Rev 04 03 2015 151 12Declaration of Conformity ...
Страница 152: ...c 12 Declaration of Conformity 152 Rev 04 03 2015 b ...
Страница 155: ...c 13 Index b Rev 04 03 2015 155 ...