WARNINGS:
Use only the Belmont supplied disposable. The disposable set is a single-use,
sterile, non-pyrogenic fluid path.
Do not change or alter the disposable. Use extreme caution when handling the
disposable; do not expose the disposable to sharp objects.
Do not use in the presence of flammable anesthetics.
Do not open the Heater Unit during or after priming.
Do not open the Heater Unit during operation, as this could damage the
disposable.
Use only crystalloid to prime the unit. Do not use blood for the initial prime.
Do not attempt to prime unit or subject the disposable to flow outside of the
Heater Unit, as the disposable will be damaged.
The patient line must be completely free of air before fluid is administered.
When administering blood products, be sure to use the administration set with
a built-in filter.
Do not use for administration of cryo-precipitate, or granulocyte suspensions.
Use only anticoagulated blood products.
Do not mix lactated Ringer’s or other solutions containing calcium with citrated
blood products.
Practice standard hospital policy when handling blood products. Treat blood as
if it were infected and clean up all spills immediately.
6