OPERATING PARAMETERS
Flow Rate
Input temperature 20°C: Max 80 ml/min
Input temperature 10°C: Max 50 ml/min
Output Temperature:
38° ± 2°C
Prime Volume
4 ml without tubing,
7.5 ml with valves and tubing at the input and output
OPERATING PANEL
Control Keys
(On Battery Housing
)
On/off
Switch – Turn battery ON/OFF
LED
(On Battery Housing
)
BLUE: Solid – Battery power is ON
RED: Flashing once at start-up – Heater Unit is working
RED: Flashing alternate with BLUE flashing - Alarm
LED
(On Heater Unit)
GREEN: Flashing with each heating cycle.
SAFETY AND MONTORING
Infusate Temperature
Heater power shut off if fluid temperature exceeds
specified limits
Independent Safety Circuit
Shut off the heater if temperature exceeds a limiting value
Alarm Condition
CHECK FOR AIR
NO HEAT
OVER TEMPERATURE
PROBE FAULT
DISPOSABLE SET
Disposable Set
Sterile, Non-Pyrogenic Fluid Path, Single-Patient Use
Only
Sterilization Method
Ethylene Oxide
Sterility Assurance Level
Greater than or equal to 10
-6
20