Operating Steps
Detailed Information
Establish Power to the Heater Unit
Depress the ON/OFF switch, on
the battery housing, to supply power
to the Heater Unit.
Blue LED turns ON and Red LED
flashes once, if the Heater Unit is
functioning normally. If not, return
for service.
Initiate Infusion
PATIENT
MALE LUER
FEMALE LUER
CONNECT TO
FLUID
BED
SHEET
CLAMP
•
Select an appropriate catheter
size for the intended flow rate.
•
Using aseptic technique, make
patient connection without
entrapping air.
•
Adjust the roller clamp to the
desired flow rate.
•
Secure the system in a
horizontal position, on a flat
surface near the patient
using
bed sheet clamp ensuring that
there is no strain on the patient
line.
•
If a longer administration set is
desired, it should be no longer
than 6 inches for optimal
thermal transfer. Use of longer
set will result in some line
cooling from heater to patient.
•
The preferred position, of the
Heater Unit, is at or below the
level of the insertion site.
9