Operating Steps
Detailed Information
Maintain Infusion
There are no adjustments on the Heater Unit.
You may see some condensation around the outer-edge of
the Heater Unit after running for an extended period of time.
This is normal and will not interfere with the system.
If withdrawing fluid from distal end of disposable set via a
syringe, do not exceed a rate of withdrawal of 100 ml/min.
•
Observe the LED, on the
battery housing. The LED is
BLUE when the battery power
is ON.
•
The GREEN LED, on the
Heater Unit, flashes with each
heating cycle.
•
If there is an alarm, the BLUE
and RED LEDs flash
alternately, respond to and
correct system alarms as
needed.
End of Procedure
•
Clamp roller clamp on the
administration set off.
•
Clamp the input end of
disposable set off using the
blue slide clamp.
•
Turn Heater Unit off by
pressing the ON/OFF switch
on the battery housing.
•
Open the Heater Unit and
remove disposable set.
•
Dispose of disposable set using
hospital practices for bio-
hazardous materials.
•
Remove the battery by pushing
the tab holding the battery, in
the housing, up and pull battery
tab.
•
Recharge the battery.
10