ELECTROMAGNETIC COMPATIBILITY
The Belmont
®
buddy lite
is intended for use in the electromagnetic environment
specified below. You should assure that it is used in such environment.
Test
Compliance
Electromagnetic environment -
guidance
RF emissions,
CISPR 11
Group 1, Class B
The
buddy lite
is a battery operated
Blood/Fluid Warmer. RF emission is
negligible at any single frequency.
RS103 Radiated E-
Field Susceptibility
RS103 of MIL-STD
461F for
(Aircraft Internal)
The
buddy lite
was tested and passed in
accordance with the radiated
susceptibility requirements to RS103 of
MIL-STD 461F for Aircraft Internal
applications.
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Radiated RF
IEC 61000-4-3
3 V/m over the
frequency range of 80
MHz to 2.5 GHz
Radiated RF frequency is negligible at any
single frequency.
Power Frequency
Magnetic Field
Immunity
IEC 61000-4-8
3 A/m
If interference occurs, it may be
necessary to position the buddy lite
further from sources of power frequency
magnetic field.
18