Temperature Calibration Procedure
Introduction
6
If the measured and set temperatures are different, note how many degrees different they are
and adjust the set temperature up or down that number of degrees.
For example:
If the required sample
temperature is
And the measured buffer/
water temperature is
Set the temperature
to
4°C
6°C
2°C
5°C
4°C