PN B38322AA
B-1
APPENDIX B
Temperature Calibration Procedure
Introduction
The Avanti JXN series centrifuge specification for temperature control is
±
2°C. The following
procedure is provided for those cases in which temperature control within
±
1°C is required.
In this procedure, you will measure the temperature of your sample after equilibrating the rotor
temperature for a minimum of 1 hour. Based on the test results, you can then adjust the centrifuge
temperature setting to bring your sample as close as possible to the required temperature.
Calibrating Temperature
1
For runs at other than ambient temperature, precool or prewarm the rotor and test samples
to the required temperature.
The length of precooling/prewarming time depends on how different the rotor starting
temperature is from the required run temperature.
2
Install the rotor with adapters, if applicable, and tubes or bottles filled with sample buffer or
water (if above 2°C).
3
Set the speed to 2000 RPM, the time to
HOLD
, and the temperature to the required run
temperature.
a.
Start the run.
4
After 30 minutes, set the required run speed, and allow the system to run for at least 1 hour.
NOTE
The system must run for at least 1 hour before the temperature of the buffer or water is
measured
. The length of time required for equilibration depends on the rotor and centrifuge starting
temperatures and the ambient room temperature.
5
After at least 1 hour, measure the temperature of the buffer or water using a thermometer or
thermocouple.
Содержание Avanti JXN-26
Страница 4: ...PN B38322AA iv Revision History...
Страница 21: ...xxi Contents Abbreviations Index Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty Related Documents...
Страница 30: ...PN B38322AA xxx Introduction Symbols Symbols...
Страница 32: ...PN B38322AA xxxii Introduction Graphics...
Страница 44: ...PN B38322AA 1 12 Description Available Rotors...
Страница 54: ...PN B38322AA 2 10 The Touch Screen Interface About Your Instrument...
Страница 154: ...PN B38322AA 6 60 Functional Pages Zonal Operation Page...
Страница 168: ...PN B38322AA 7 14 Maintenance and Troubleshooting Supply List...
Страница 176: ...PN B38322AA A 8 Preinstallation Requirements Using J2 Series Rotors in the Avanti JXN Series...
Страница 194: ...PN B38322AA D 12 Third Party Contributions and Legal Notices Contributions and Legal Notices...
Страница 196: ...PN B38322AA E 2 HEPA Filter Replacement Instructions for the Avanti JXN Series Centrifuges HEPA Filter Replacement...
Страница 198: ...PN B38322AA Abbreviations 2 Abbreviations...
Страница 206: ...Index 8 Index...
Страница 208: ...PN B38322AA Warranty 2 Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty...
Страница 209: ......