PN B38322AA
7-11
Maintenance and Troubleshooting
JCF-Z Rotor Identification
7
3
Repeat step
on the right side
The second latch engages.
JCF-Z Rotor Identification
The rotor identification system can, under two specific conditions, misidentify the JCF-Z
continuous flow/zonal rotor. If the rotor is misidentified, the the run will be stopped. Follow the
steps below to resolve the issue.
NOTE
Misidentification can be prevented by returning your JCF-Z rotor to the Beckman Coulter factory to
have magnets added to the rotor body. The magnets ensure correct identification. Call your local
Beckman Coulter Representative
*
for more information.
JCF-Z rotors manufactured after March, 1997, have factory-installed magnets.
1
On the first few uses of a new JCF-Z rotor, or in an older JCF-Z rotor with newly replaced bearings
, run
the rotor from 0 to 5000 RPM and back to 0 RPM three times.
This procedure will distribute lubricant around the bearings in the rotating seal assembly,
reducing drag.
2
In older JCF-Z rotors with bearings that have not been replaced recently
, follow the same procedure
as for a new JCF-Z rotor (perform three runs from 0 to 5000 RPM and back to 0 RPM) to ensure
that the bearings are properly lubricated.
If misidentification recurs after the third run, when the bearings are lubricated, this indicates
that the bearings are worn and need to be replaced.
a.
Replace the bearings following the instructions in the JCF-Z manual (JCFZ-IM).
NOTE
If the JCF-Z rotor is misidentified, the run will be stopped.
Storage and Transportation
To ensure that the instrument does not get damaged, contact your Beckman Coulter Representative
for specific instructions and/or assistance in preparing the equipment for transport or long-term
storage.
*
In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office or visit us at
www.beckmancoulter.com.
Содержание Avanti JXN-26
Страница 4: ...PN B38322AA iv Revision History...
Страница 21: ...xxi Contents Abbreviations Index Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty Related Documents...
Страница 30: ...PN B38322AA xxx Introduction Symbols Symbols...
Страница 32: ...PN B38322AA xxxii Introduction Graphics...
Страница 44: ...PN B38322AA 1 12 Description Available Rotors...
Страница 54: ...PN B38322AA 2 10 The Touch Screen Interface About Your Instrument...
Страница 154: ...PN B38322AA 6 60 Functional Pages Zonal Operation Page...
Страница 168: ...PN B38322AA 7 14 Maintenance and Troubleshooting Supply List...
Страница 176: ...PN B38322AA A 8 Preinstallation Requirements Using J2 Series Rotors in the Avanti JXN Series...
Страница 194: ...PN B38322AA D 12 Third Party Contributions and Legal Notices Contributions and Legal Notices...
Страница 196: ...PN B38322AA E 2 HEPA Filter Replacement Instructions for the Avanti JXN Series Centrifuges HEPA Filter Replacement...
Страница 198: ...PN B38322AA Abbreviations 2 Abbreviations...
Страница 206: ...Index 8 Index...
Страница 208: ...PN B38322AA Warranty 2 Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty...
Страница 209: ......