PN B38322AA
7-6
Maintenance and Troubleshooting
Diagnostics/User Messages
1
Remove the centrifuge front panel.
Accessing the Rotor in Case of Power Failure
for instructions.
2
Unscrew the filler cap. Set the cap aside.
3
Add oil (PN 392760) up to the maximum fill line. Pour the oil through a funnel if necessary.
4
Screw the filler cap back on tightly.
5
Replace the front panel. See
Replacing the Front Panel on the Centrifuge
for instructions.
Diagnostics/User Messages
When a condition arises that requires operator attention, the header bar turns yellow or red. A page
displays the diagnostic message. User messages communicate information about the centrifuge or
alert you to an abnormal condition. For a list of the possible malfunctions and their corrective
actions, see
Accessing the Rotor in Case of Power Failure
WARNING
Any maintenance procedure requiring removal of a panel exposes the operator to
the possibility of electrical shock and/or mechanical injury. Therefore, turn the
power off (O) and disconnect the instrument from the main power source by
removing the Mains (power) plug from the outlet receptacle, and refer such
maintenance to qualified service personnel.
WARNING
Before performing this procedure, verify that the rotor is not spinning by listening
carefully for any noise coming from the chamber. Proceed only if the instrument
is quiet. Never attempt to override the door interlock system while the rotor is
spinning.
If the facility power fails only momentarily, the centrifuge will resume operation when the power
is restored and the rotor will return to the set speed. In the event of an extended power failure,
Содержание Avanti JXN-26
Страница 4: ...PN B38322AA iv Revision History...
Страница 21: ...xxi Contents Abbreviations Index Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty Related Documents...
Страница 30: ...PN B38322AA xxx Introduction Symbols Symbols...
Страница 32: ...PN B38322AA xxxii Introduction Graphics...
Страница 44: ...PN B38322AA 1 12 Description Available Rotors...
Страница 54: ...PN B38322AA 2 10 The Touch Screen Interface About Your Instrument...
Страница 154: ...PN B38322AA 6 60 Functional Pages Zonal Operation Page...
Страница 168: ...PN B38322AA 7 14 Maintenance and Troubleshooting Supply List...
Страница 176: ...PN B38322AA A 8 Preinstallation Requirements Using J2 Series Rotors in the Avanti JXN Series...
Страница 194: ...PN B38322AA D 12 Third Party Contributions and Legal Notices Contributions and Legal Notices...
Страница 196: ...PN B38322AA E 2 HEPA Filter Replacement Instructions for the Avanti JXN Series Centrifuges HEPA Filter Replacement...
Страница 198: ...PN B38322AA Abbreviations 2 Abbreviations...
Страница 206: ...Index 8 Index...
Страница 208: ...PN B38322AA Warranty 2 Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty...
Страница 209: ......