13-2
User Guide
CHAPTER 13
Troubleshooting
Measurement does
not start (indicators
remain red)
Verify that you are using the cor-
rect probe, and that the
Start
button (or foot-pedal) has been
pressed.
Verify that sufficient ultrasound
coupling gel has been applied,
and to the correct area. Check
that the probe is in good contact
with the patient’s skin.
Validate the system by executing
the System Quality Verification
procedure.
Check that the probe is properly
connected and the connector
screws tightened.
Abort and restart the measure-
ment if necessary.
The Acoustic Contact
Indicator shows
green but the Acous-
tic Signal Acquisition
Indicator remains red
Validate the system by executing
the System Quality Verification
procedure.
Check that the probe is properly
positioned for the specific site
being measured.
If the patient’s soft tissue is
extremely thick at the site, mea-
surement may be obstructed. In
particular, patients with a Body
Mass Index (BMI) greater than
35 may exhibit this problem. In
this case additional pressure to
bring the probe closer to the
bone may solve the problem.
Result of System
Quality Verification is
outside phantom
range
Call Sunlight Customer Support
Problem
Possible Solution
Содержание Sunlight Omnisense 7000S
Страница 4: ...iv ...
Страница 12: ...xii ...
Страница 18: ...xviii ...
Страница 22: ...xxii ...
Страница 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Страница 52: ...2 20 User Guide CHAPTER 2 Essential Prescribing Information Figure 2 2 Example of a Measurement Report ...
Страница 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Страница 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Страница 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Страница 174: ...9 12 User Guide CHAPTER 9 Interpreting the Results of a Measurement Figure 9 8 Example of a Measurement Report ...
Страница 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Страница 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Страница 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...
Страница 278: ...17 14 User Guide CHAPTER 17 Research Mode ...