Omnisense 7000S/8000S
8-3
Readying Omnisense for an SOS Measurement
8.2 Readying Omnisense for an SOS Measurement
The procedure below describes your interaction with the
Omnisense software in preparation for an SOS measurement.
This is done first, so that you are free to concentrate on the scan
site during the measurement itself.
Readying Omnisense
S
TEP
1 Select measurement type
Setting the
measurement type
If you have not already done so, select “Single Site
Measurement” in the
Measurement Type
selection
box.
Refer to Figure 8.1. The other measurement
type is
System Quality Verification
, described in
Chapter 6, "System Quality Verification‚" page 6-1.
S
TEP
2 Identify the patient
All SOS measurements must be conducted with ref-
erence to a specific patient known to the Omnisense
software. Please refer to Figure 8.1.
Open the patient file for whom the SOS measure-
ment is to be conducted. Click on the
Open Patient
button at the upper right of the patient information
section of the screen, or choose
Open Patient
from
the Patient menu. Select the patient from the list
which is displayed — the patient is now open. For a
new patient, create a new patient record. For more
information on opening and creating patient files,
see Chapter 7, "Patient Files‚" page 7-1.
After a patient file has been opened, the
Measure-
ment
button on the toolbar becomes active (next
step).
Содержание Sunlight Omnisense 7000S
Страница 4: ...iv ...
Страница 12: ...xii ...
Страница 18: ...xviii ...
Страница 22: ...xxii ...
Страница 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Страница 52: ...2 20 User Guide CHAPTER 2 Essential Prescribing Information Figure 2 2 Example of a Measurement Report ...
Страница 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Страница 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Страница 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Страница 174: ...9 12 User Guide CHAPTER 9 Interpreting the Results of a Measurement Figure 9 8 Example of a Measurement Report ...
Страница 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Страница 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Страница 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...
Страница 278: ...17 14 User Guide CHAPTER 17 Research Mode ...