Omnisense 7000S/8000S
12-1
CHAPTER 12
Cleaning and Disinfecting
Omnisense
12.1 Cleaning and Disinfection
General Instructions
The Omnisense probe should not be used on subjects with
breached skin or open sores on the skin area that comes into
contact with the probe.
The Omnisense probe requires proper cleaning and disinfecting
between each patient use. Doing so may help prevent any
potential transmission of infection between patients. In case of
using probe caps, disinfection is not required.
Protective gloves are recommended for cleaning and disinfect-
ing procedures.
Soiled materials should be disposed of in appropriate waste
receptacles.
The Omnisense Device (7000S and 8000S)
CLEANING
Cleaning should be done with the probes connected, to make
sure that no liquid gets into the probe sockets. No disinfection is
required for the Omnisense unit since there are no patient con-
tact surfaces.
The following steps should be followed for cleaning the
Omnisense device:
1.
Use a soft cloth, lightly moistened with liquid detergent or
mild soap, to wipe down Omnisense’s device surface.
Содержание Sunlight Omnisense 7000S
Страница 4: ...iv ...
Страница 12: ...xii ...
Страница 18: ...xviii ...
Страница 22: ...xxii ...
Страница 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Страница 52: ...2 20 User Guide CHAPTER 2 Essential Prescribing Information Figure 2 2 Example of a Measurement Report ...
Страница 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Страница 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Страница 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Страница 174: ...9 12 User Guide CHAPTER 9 Interpreting the Results of a Measurement Figure 9 8 Example of a Measurement Report ...
Страница 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Страница 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Страница 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...
Страница 278: ...17 14 User Guide CHAPTER 17 Research Mode ...