Performing System Quality Verification
If the results of measurement cycles are not being
accepted by Omnisense, check that no air bubbles are
being trapped between the probe and the phantom, and
that sufficient ultrasound gel has been applied. If results
are still not being accepted, your probe may be defective.
Contact
BeamMed
Customer Support.
S
TEP
11 Go to Result
Click on the
Result
button and the following is dis-
played.
Figure 6.7
Entering the expected phantom SOS value
Enter the number noted in Step 5 and click OK.
S
TEP
12 Check the result
The result is displayed graphically along with
results from recent System Quality Verifications, as
illustrated in Figure 6.8.
S
TEP
13 Print Result
If desired, the SQV result can be printed at this
point, by pressing the Print button on the toolbar,
seen at left.
SQV Measurement>Result
Enter value and OK
Содержание Sunlight Omnisense 7000S
Страница 4: ...iv ...
Страница 12: ...xii ...
Страница 18: ...xviii ...
Страница 22: ...xxii ...
Страница 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Страница 52: ...2 20 User Guide CHAPTER 2 Essential Prescribing Information Figure 2 2 Example of a Measurement Report ...
Страница 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Страница 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Страница 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Страница 174: ...9 12 User Guide CHAPTER 9 Interpreting the Results of a Measurement Figure 9 8 Example of a Measurement Report ...
Страница 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Страница 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Страница 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...
Страница 278: ...17 14 User Guide CHAPTER 17 Research Mode ...