6-8
User Guide
CHAPTER 6
System Quality Verification
the current one. Reposition the phantom on the
probe and perform the measurement steps 6-9 twice
more. Add additional Parker Gel if required. After
three cycles have been completed, the
Result
button
at the lower right of the screen should become
active.
Figure 6.6
SQV measurement cycle completed
If the
Result
button does not become active after
three cycles, this indicates that one or more of the
measurement cycles yielded statistically inconsis-
tent results and is not usable. Repeat for another
measurement cycle. Omnisense will allow up to
five cycles; if three consistent measurement cycles
have not been obtained after five cycles the mea-
surement is aborted and the counter cycles back to
1. In this case you must begin again and attempt
three new cycles.
Содержание Sunlight Omnisense 7000S
Страница 4: ...iv ...
Страница 12: ...xii ...
Страница 18: ...xviii ...
Страница 22: ...xxii ...
Страница 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Страница 52: ...2 20 User Guide CHAPTER 2 Essential Prescribing Information Figure 2 2 Example of a Measurement Report ...
Страница 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Страница 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Страница 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Страница 174: ...9 12 User Guide CHAPTER 9 Interpreting the Results of a Measurement Figure 9 8 Example of a Measurement Report ...
Страница 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Страница 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Страница 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...
Страница 278: ...17 14 User Guide CHAPTER 17 Research Mode ...