Medical Gas Central Alarm - Medipoint 125 Digital Alarm

4233500107.03

Medipoint 125 consists of: -

No. Description

Control Fascia

1

Power on LED.

2

Temperary mute button.

3

System Fault LED.

4

Test button.

LCD Touch Screen

5

Information screen.

6

Data logging screen.

7

Settup Menus.

8

Alarm service location.

9

Emergency battery supply status.

10

Gas service type

11

Gas service colour identification bar

12

Gas service status area

13

Indication that date/time is not set

14

Mute activated

1. 

General Information.

1.1 

Introduction.

The  BEACONMEDÆS  Medipoint  125  digital  (MP125D) 

medical gas cental alarm is suitable for both the United 

Kingdom  and  International  markets,  and  fully  satisfies 

the  requirements  the  requirements  of  the  HTM02-01, 

HTM2022 and ISO7396-1 applications. . 

The Central Alarm is defined in ISO 7396-1 as ‘‘Operating 

Alarms’.  The  intention  of  the  alarm  is  to  indicate  to 

technical  staff  that  it  is  necessary  to  replenish  the  gas 

supply or to correct a malfunction within the Medical Gas 

Pipeline System that requires an immediate response.

Each alarm panel requires a standard AC electrical power 

supply of 115/230 volts ±10% at 50 or 60 Hz, and fused 

at 3 Amps.

The central alarm can be programmed from on screen 

menus to provide a fully flexible alarm system.  Allowing 

the panels to be adapted to suit any hospital’s medical 

gas  monitoring  requirements  (see  section  1.12  for 

programming details).

An individual central alarm panel has 5 inputs or up to 

1

4

2

3

6 columns for displaying as a repeater panel consisting 

of ‘normal’ and 4 ‘fault’ conditions.  These columns can 

be setup to monitor either up to 5 medical gas source 

equipment,  or  up  to  20  individual  point  alarms,  or  a 

combination of the 2.

One central alarm panel is designated the master panel 

for  control  purposes.  Any  number  of  panels  may  be 

programmed as a ‘central acknowledge panel’ and cause 

all flashing displays on the system to change to steady 

when  the  MUTE  switch  on  the  central  acknowledge 

panel is operated. 

Each  central  alarm  panel  incorporates  flashing  alarm 

displays,  a  POWER  ON  indication,  a  SYSTEM  ALARM 

indication,  MUTE  and  TEST  switches  and  an  audible 

warning  programmed  to  operate  depending  upon  the 

9

12

10

11

5

6

7

8

Figure 1 - 

Medipoint 125 Digital Alarm

 (MP125D).

13

14

Safety Notice

Persons undertaking any installation and/or maintenance 

must be fully trained in specialist work of this nature.

The “PERMIT TO WORK” procedure must be adhered to 

for all installations once commissioned.

The alarm is designed and built in accordance with HTM 

02-01 and ISO 7396-1 regulations and therefore should 

be installed as such.

Содержание 8102371401

Страница 1: ...Operation and Maintenance Instructions Medical Gas Central Alarm Medipoint 125 Digital Alarm Part number 4233500107 Revision 03 Jan 2021...

Страница 2: ...hone 44 0 1246 474242 Email gbn info beaconmedaes com Website Contacts www beaconmedaes com BeaconMed s reserves the right to make changes and improvements to update products sold previously without n...

Страница 3: ...in screen 3 6 Exit Setup Screen 3 7 Confi guration Menu Map 4 0 Commissioning 4 1 Alarm panel test 4 2 Checking the slave relay output 4 3 Checking the remote audible warning 4 4 Medical gas alarm fau...

Страница 4: ...ckaged and stored in controlled conditions Environmental Operating Conditions Adverse environmental conditions and harsh abrasives or chemicals may cause damage to the unit Environmental Protection Di...

Страница 5: ...rammed from on screen menus to provide a fully flexible alarm system Allowing the panels to be adapted to suit any hospital s medical gas monitoring requirements see section 1 12 for programming detai...

Страница 6: ...nt of any defect within the monitored functions the SYSTEM ALARM LED illuminates flashing accompanied by the audible warning and the defective gas service alarm display status bar illuminates An inter...

Страница 7: ...operty and shall not without our permission be altered ATLAS COPCO STANDARD Class Tolerances if not indicated according to Material Treatment Name Scale Family Drawing owner Secrecy Class 1102 K Compa...

Страница 8: ...p to four fault conditions yellow or red flashing alarm conditions to show the gas service status All alarm conditions are illuminated by a flashing display and accompanied by an audible warning Each...

Страница 9: ...on these PCB s which are interconnected by means of a multi way ribbon cable and polarised connector see figure 2 3 The area central alarm internal electrical installation complies with all relevant...

Страница 10: ...cular displayed alarm condition only This facility is designed for use when the plant or pipeline is shut off for a prolonged period On returning the pipeline pressure to NORMAL this feature automatic...

Страница 11: ...nnections have 5 terminals each See table 2 for details for standard options Each gas service and fault legend is also fully customisable through the on screen setup menus TABLE 2 GAS SERVICE CONNECTI...

Страница 12: ...uired by the contract specification Feed cables into the box leaving 400mm to 75 150 105 105 7 260 5 TOP 40 40 40 40 17 20 KNOCKOUT SEGMENTS 200 17 45 2 0 KNOCKOUT SEGMENTS BACKBOX SURFACE Figure 10 B...

Страница 13: ...Power supply PCB installation No Description QTY 1 Back Box 1st fix supply N A 2 Power supply PCB 2nd fix supplied loose 1 3 PCB Spacer 2nd fix supplied loose 4 4 M3 Nut 2nd fix supplied loose 4 5 M3...

Страница 14: ...nections No Description QTY 1 Earth cable 2nd fix supplied assembled 1 2 M3xNut 2nd fix supplied loose 4 3 M3 washer 2nd fix supplied loose 4 Typical earth stud assembly 2 2 3 1 Alarm Door Earth Stud...

Страница 15: ...igure 15 Alarm data connection MP125D 1 S1 ON IN OUT A B A B S MP125D 2 S1 OFF IN OUT A B A B S MP125D 3 S1 ON IN OUT A B A B S Note S terminal is for connecting the screen cable 2 2 6 Remote audible...

Страница 16: ...mmissionedimmediately disconnect the standby battery Connect electrical power supply wires to electrical power supply plug terminals E N and L 1 2 Figure 21 Alarm door security fastener No Description...

Страница 17: ...aracters Icons Icon Description Enter Key TAB key Moves to next field and highlights text Highlighted text will be overwritten on next key press Tick key Apply update and takes you to the main screen...

Страница 18: ...e Alarm setup 2 screen or press the Exit Icon to go back to the main operating screen 3 4 Alarm Setup Screen 1 Figure 26 The following steps describes the setup process for the alarm setup screen 1 3...

Страница 19: ...e new numerical password Press the TAB key or press the next field to go to the next field Confirm Use the num pad to enter the password again Press the ENTER key or press anywhere on the screen away...

Страница 20: ...setup process for the main gas service setup screen Figure 27 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Tak...

Страница 21: ...k to set as inputs tick to set as remote transmission Press to toggle between the options 3 5 6 Location Special Characters Icons Description Exit Icon Takes you back to the main screen O2 Oxygen Gas...

Страница 22: ...elds will be editable 3 5 7 1 Standard gas options These fields allow for selecting the source equipment type available for the standard gas type selected Press on the matching equipment type to auto...

Страница 23: ...etup screen Figure 33 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Ico...

Страница 24: ...key press Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Icon Takes you back to the main screen Reject Icon Rejects changes Accept Icon Accepts...

Страница 25: ...6 shows examples of all status flashing See table 6 in section 5 Operation of the Medipoint 125 Digital alarm for troubleshooting Figure 36 Alarm Test Button example 6 gas 4 2 Checking the slave relay...

Страница 26: ...e table 2 section 1 8 for full list of standard gas service fault legends Check that the MUTE switch operates satisfactorily Press the audible alarm will be silenced and the mute icon will show on scr...

Страница 27: ...ED is flashing red The audible alarm is active See figure 39 fro example Figure 39 Example Gas service 1 connection fault 4 5 3 Input Reconnect Reconnect the input plug to gas channel System will retu...

Страница 28: ...d and the alarm condition remains the audible warning will re sound after 15 mins and a further MUTE selection will be required Should a further alarm condition occur after the panel has indicated a f...

Страница 29: ...mple shows gas service fault 4 Power LED Green On System Fault Black Off Gas Status Bar Red Flashing Audible Active Audible Mute Gas service fault flashing Mute activated Audible Silenced for 15 minut...

Страница 30: ...lank normal Yellow flashing fault Fault status 3 Blank Fault status 4 Blank normal Red flashing fault Manifold Control Systems Normal status Green normal blank during fault Fault status 1 Blank normal...

Страница 31: ...to the main screen automatically if a new alarm condition becomes active or after one minute if the alarm is not manually exited from the information screen 5 2 2 Data Logging Screen The data logging...

Страница 32: ...s distributors will be able to supply similar contracts in other areas CAUTION It is essential that only genuine BeaconMed ssparepartsarefittedduringmaintenance WARNING Obtain a work permit or equival...

Страница 33: ...33 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45a Alarm panel assembly details...

Страница 34: ...34 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45b Alarm panel assembly details...

Страница 35: ...e 5 year maintenance replace the LCD screen and backup butteries See maintenance procedures in 6 1 and figures 45a 45b for details 8 RECOMMENDED SPARES Table 8 details the list of recommended spares t...

Страница 36: ...2460 Atlas Copco Medical Ltd Telford Crescent Staveley Derbyshire S43 3PF UK Tel 44 0 1246 474242 www beaconmedaes com Atlas Copco Airpower n v Boomsesteenweg 957 2610 Wilrijk Belgium...

Отзывы: